Embolic protection filters for carotid stenting: differences in flow obstruction depending on filter construction.

机译:用于颈动脉支架置入术的栓塞保护过滤器:流阻的差异取决于过滤器的结构。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To investigate the pressure gradient and degree of flow reduction associated with embolus protection filters for carotid stenting in an in vitro experiment. METHODS: Three filter devices with a perforated membrane design and one wire mesh type filter were tested. At a pressure of 70 mmHg, the flow reduction and pressure gradient were measured in a 5-mm tube using blood-mimicking fluid. RESULTS: The pressure gradient in the wire mesh filter was 1.65+/-0.49 mmHg (95% CI 1.32 to 1.86). The mean pressure gradient in the perforated membrane filters was 6.88+/-2.62 mmHg (95% CI 6.22 to 7.55, p<0.0001). There was also a significant correlation between pressure gradient and flow reduction (r=-0.77, p<0.01). CONCLUSION: Embolic protection filters cause a pressure gradient and obstruct blood flow. This effect is marked in perforated membrane filters and almost absent in the wire mesh filter.
机译:目的:在体外实验中研究与用于颈动脉支架置入术的栓塞保护过滤器相关的压力梯度和减少流量的程度。方法:测试了三种带孔膜设计的过滤器设备和一个丝网式过滤器。在70 mmHg的压力下,使用模拟血液的液体在5毫米管中测量流量减少和压力梯度。结果:丝网过滤器中的压力梯度为1.65 +/- 0.49 mmHg(95%CI为1.32至1.86)。多孔膜过滤器中的平均压力梯度为6.88 +/- 2.62 mmHg(95%CI 6.22至7.55,p <0.0001)。压力梯度和流量减少之间也存在显着相关性(r = -0.77,p <0.01)。结论:栓塞防护过滤器会引起压力梯度并阻碍血液流动。这种效果在多孔膜过滤器中很明显,而在丝网过滤器中几乎没有。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号