首页> 外文期刊>Journal of Electroanalytical Chemistry: An International Journal Devoted to All Aspects of Electrode Kinetics, Interfacial Structure, Properties of Electrolytes, Colloid and Biological Electrochemistry >The highly sensitive detection of catecholamines using a microfluidic device integrated with an enzyme-modified pre-reactor for interferent elimination and an interdigitated array electrode
【24h】

The highly sensitive detection of catecholamines using a microfluidic device integrated with an enzyme-modified pre-reactor for interferent elimination and an interdigitated array electrode

机译:使用与酶修饰的预反应器集成在一起的微流控设备对儿茶酚胺进行高灵敏度检测,以消除干扰,并使用叉指阵列电极

获取原文
获取原文并翻译 | 示例
           

摘要

We developed a microfluidic device for the continuous monitoring of catecholamine levels and studied the performance of two components of the device, namely an enzyme-modified pre-reactor and an interdigitated array (IDA) electrode. The enzyme-modified pre-reactor and IDA electrode were integrated on a chip. The integrated pre-reactor with a 3D structure, consisted of rectangular-shaped micropillars immobilized with ascorbate oxidase, and was used for the enzymatic removal of the electroactive interferent, L-ascorbic acid (AA). It exhibited elimination efficiencies of 99.7% and 98% for AA concentrations of 10 and 100 mu M, respectively. We used a carbon-based IDA electrode (2 mu m line and space) fabricated in a microflow channel for the highly sensitive detection of CAs. Redox cycling at the IDA electrode and an increase in the mass transport to the electrode contributed to the highly sensitive detection of dopamine (DA) and adrenaline (AD). The detection limits of the device for DA and AD standard solution in the absence of AA were 108 and 440 pM, respectively. (c) 2005 Elsevier B.V. All rights reserved.
机译:我们开发了一种用于连续监测儿茶酚胺水平的微流控设备,并研究了该设备的两个组件的性能,即酶修饰的预反应器和叉指阵列(IDA)电极。酶修饰的预反应器和IDA电极集成在芯片上。集成的具有3D结构的预反应器,由固定有抗坏血酸氧化酶的矩形微柱组成,用于酶促去除电活性干扰物L-抗坏血酸(AA)。对于10和100μM的AA浓度,其去除效率分别为99.7%和98%。我们使用微流道中制造的碳基IDA电极(行距和间隔为2μm)对CA进行高度灵敏的检测。 IDA电极上的氧化还原循环和向电极的质量传输增加有助于多巴胺(DA)和肾上腺素(AD)的高灵敏度检测。在没有AA的情况下,用于DA和AD标准溶液的设备的检测限分别为108 pM和440 pM。 (c)2005 Elsevier B.V.保留所有权利。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号