BACKGROUND: A 35-year-old female presented with complaints of tiredness, shortness of breath and recent onset of chest pain. The cardiac troponin I (cTnI) concentration was elevated at 6.4 mug/L (Abbott Architect reference value: <0.03 mug/L) with a normal ECG. Physical examination, radiological imaging and routine laboratory investigations did not provide an explanation for the elevated cTnI concentration. METHODS: Use of different troponin assays, dilutions with assay diluent, addition of mouse serum, polyethylene glycol (PEG) precipitation and gel filtration chromatography (GFC) were used to investigate possible interference. RESULTS: Troponin concentrations were below the level of detection when measured using all other troponin immunoassays. Serial dilutions of sample and addition of mouse serum did not alter the results. However, PEG precipitation and GFC showed the presence of a high molecular weight immunoreactive protein. Using GFC and protein-A IgG precipitation, the interference could be identified as a macrocomplex containing IgG and (fragments of) cTnI. CONCLUSIONS: We report a false positive cTnI result caused by a true macrotroponin, containing IgG and (fragments of) cTnI. This macrotroponin was only immunoreactive in the Abbott Architect cTnI immunoassay. Clinicians should be aware of analytical interference when troponin results are constantly elevated in the absence of (non)coronary causes of an increased troponin.
机译:背景:一名35岁的女性表现出疲倦,呼吸急促和最近发作的胸痛。心电图正常时,心肌肌钙蛋白I(cTnI)浓度升高至6.4杯/升(Abbott Architect参考值:<0.03杯/升)。体格检查,放射影像学检查和常规实验室检查未提供cTnI浓度升高的解释。方法:使用不同的肌钙蛋白测定法,用测定稀释剂稀释,添加小鼠血清,聚乙二醇(PEG)沉淀和凝胶过滤色谱法(GFC)用于研究可能的干扰。结果:使用所有其他肌钙蛋白免疫测定法测定的肌钙蛋白浓度低于检测水平。样品的连续稀释和添加小鼠血清不会改变结果。但是,PEG沉淀和GFC显示存在高分子量免疫反应蛋白。使用GFC和蛋白A IgG沉淀,可以将干扰物鉴定为包含IgG和cTnI(片段)的大复合体。结论:我们报告了由真正的大肌钙蛋白引起的cTnI假阳性结果,其中含有IgG和cTnI(片段)。该大肌钙蛋白仅在Abbott Architect cTnI免疫测定中具有免疫反应性。当肌钙蛋白结果持续升高而没有(非)冠心病肌钙蛋白升高的原因时,临床医生应意识到分析干扰。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号