European multicenter analytical evaluation of the Abbott ARCHITECT STAT high sensitive troponin i immunoassay

机译:欧培ARCHITECT STAT高灵敏度肌钙蛋白I免疫测定的欧洲多中心分析评估

获取原文
获取原文并翻译 | 示例
           

摘要

Background: International recommendations highlight the superior value of cardiac troponins (cTns) for early diagnosis of myocardial infarction along with analytical requirements of improved precision and detectability. In this multicenter study, we investigated the analytical performance of a new high sensitive cardiac troponin I (hs-cTnI) assay and its 99th percentile upper reference limit (URL).Methods: Laboratories from nine European countries evaluated the ARCHITECT STAT high sensitive troponin I (hs-TnI) immunoassay on the ARCHITECT i2000SR/i1000SR immunoanalyzers. Imprecision, limit of blank (LoB), limit of detection (LoD), limit of quantitation (LoQ) linearity of dilution, interferences, sample type, method comparisons, and 99th percentile URLs were evaluated in this study.Results: Total imprecision of 3.3%-8.9%, 2.0%-3.5% and 1.5%-5.2% was determined for the low, medium and high controls, respectively. The lowest cTnI concentration corresponding to a total CV of 10% was 5.6 ng/L. Common interferences, sample dilution and carryover did not affect the hs-cTnI results. Slight, but statistically significant, differences with sample type were found. Concordance between the investigated hs-cTnI assay and contemporary cTnI assay at 99th percentile cut-off was found to be 95%. TnI was detectable in 75% and 57% of the apparently healthy population using the lower (1.1 ng/L) and upper (1.9 ng/L) limit of the LoD range provided by the ARCHITECT STAT hs-TnI package insert, respectively. The 99th percentile values were gender dependent.Conclusions: The new ARCHITECT STAT hs-TnI assay with improved analytical features meets the criteria of high sensitive Tn test and will be a valuable diagnostic tool.
机译:背景:国际建议强调了心肌肌钙蛋白(cTns)在心肌梗死早期诊断中的优越价值,以及提高精确度和可检测性的分析要求。在这项多中心研究中,我们调查了一种新的高敏感性心肌肌钙蛋白I(hs-cTnI)测定方法的分析性能及其99%的参考上限(URL)。方法:来自9个欧洲国家的实验室评估了ARCHITECT STAT高敏感性肌钙蛋白I (hs-TnI)免疫分析在ARCHITECT i2000SR / i1000SR免疫分析仪上进行。本研究评估了不精确度,空白限(LoB),检测限(LoD),稀释度定量限(LoQ)线性,干扰,样品类型,方法比较和99%的URL。结果:总不准确度为3.3对于低,中和高对照,分别测定%-8.9%,2.0%-3.5%和1.5%-5.2%。对应于总CV的10%的最低cTnI浓度为5.6 ng / L。常见的干扰,样品稀释和残留都不会影响hs-cTnI结果。发现与样品类型略有差异,但在统计学上具有统计学意义。发现在第99个百分位数临界值时,所研究的hs-cTnI分析与当代cTnI分析之间的一致性为95%。使用ARCHITECT STAT hs-TnI包装插页提供的LoD范围的下限(1.1 ng / L)和上限(1.9 ng / L),分别在75%和57%的明显健康人群中检测到TnI。结论:第99个百分位值取决于性别。结论:具有改进分析功能的新ARCHITECT STAT hs-TnI测定符合高灵敏度Tn检测标准,将成为有价值的诊断工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号