首页> 外文期刊>Clinical Biochemistry >Multi-site analytical evaluation of a chemiluminescent magnetic microparticle immunoassay (CMIA) for sirolimus on the Abbott ARCHITECT analyzer.
【24h】

Multi-site analytical evaluation of a chemiluminescent magnetic microparticle immunoassay (CMIA) for sirolimus on the Abbott ARCHITECT analyzer.

机译:在Abbott ARCHITECT分析仪上对西罗莫司进行化学发光磁性微粒免疫测定(CMIA)的多站点分析评估。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVE: This study evaluated a new chemiluminescent magnetic microparticle immunoassay (CMIA) for sirolimus on the ARCHITECT analyzer. DESIGN AND METHODS: Patient and laboratory proficiency samples were tested at three European sites and one site in the United States. RESULTS: The CMIA total %CV's were all <8% and the Limit of Quantification (LOQ) was <1.52 ng/mL across the four sites. It cross-reacts to sirolimus metabolites F4 and F5 and showed no hematocrit interference over a range of 25% to 55%. Patient specimen correlations to three LC/MS/MS methods gave R>or=0.91 at three sites and mean biases of 14%, 25% and 39%. CMIA patient specimen correlations to the Abbott IMx gave R>or=0.94 at 2 sites and mean biases of 5.4% and 6.9%. CONCLUSIONS: CMIA is a precise and sensitive immunoassay method without hematocrit interference. It correlates well to both LC/MS/MS and immunoassay results, but shows an expected positive bias to LC/MS/MS.
机译:目的:本研究在ARCHITECT分析仪上评估了西罗莫司的一种新的化学发光磁性微粒免疫测定(CMIA)。设计与方法:在欧洲的三个地点和美国的一个地点测试了患者和实验室水平的样本。结果:四个站点的CMIA总%CV均<8%,定量限(LOQ)≤1.52 ng / mL。它与西罗莫司代谢产物F4和F5发生交叉反应,在25%至55%的范围内均无血细胞比容干扰。患者样品与三种LC / MS / MS方法的相关性在三个位置给出R>或= 0.91,平均偏差为14%,25%和39%。 CMIA患者标本与雅培IMx的相关性在2个位点的R> or = 0.94,平均偏差为5.4%和6.9%。结论:CMIA是一种精确而灵敏的免疫分析方法,无血细胞比容干扰。它与LC / MS / MS和免疫测定结果均具有良好的相关性,但显示出预期的LC / MS / MS正偏倚。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号