The study aimed to assess the accuracy and reproducibility of occlusal caries depth measurements obtained from different imaging modalities. The study comprised 21 human mandibular molar teeth with occlusal caries. Teeth were imaged using film, CCD, two different cone-beam computerized tomography (CBCT) units and a microcomputer tomography (micro-CT). Thereafter, each tooth was serially sectioned, and the section with the deepest carious lesion was scanned using a high-resolution scanner. Each image set was separately viewed by three oral radiologists. Images were viewed randomly, and each set was viewed twice. Lesion depth was measured on film images using a digital caliper, on CCD and CBCT images using built-in measurement tools, on micro-CT images using the Mimics software program, and on histological images using AxioVision Rel. 4.7. Intra- and inter-rater reliabilities were assessed according to the Bland/Altman method by calculating Intraclass Correlation Coefficients (ICCs). Mean/median values obtained with intraoral systems were lower than those obtained with 3-D and histological images for all observers and both readings. Intra-observer ICC values for all observers were highest for histology and micro-CT. In addition, intra-observer ICC values were higher for histology and CBCT than for histology and intra-oral methods. Inter-observer ICC values for first and second readings were high for all observers. No differences in repeatability were found between Accuitomo and Iluma CBCT images or between intra-oral film and CCD images. Micro-CT was found to be the best imaging method for the ex vivo measurement of occlusal caries depth. In addition, both CBCT units performed similarly and better than intra-oral modalities.
机译:该研究旨在评估从不同成像方式获得的咬合龋深度测量的准确性和可重复性。这项研究包括21颗人类下颌磨牙与咬合龋。使用胶片,CCD,两个不同的锥形束计算机断层摄影(CBCT)单元和微型计算机断层摄影(micro-CT)对牙齿成像。此后,将每颗牙齿连续切片,并使用高分辨率扫描仪扫描龋齿病变最深的切片。每个图像集由三位口腔放射科医生分别查看。随机查看图像,每组查看两次。使用数字卡尺在胶片图像上,使用内置测量工具在CCD和CBCT图像上,在使用Mimics软件程序的micro-CT图像上以及在使用AxioVision Rel的组织学图像上测量病变深度。 4.7。根据Bland / Altman方法,通过计算类内相关系数(ICC)来评估评定者内部和评定者之间的可靠性。对于所有观察者和两个读数,通过口内系统获得的均值/中值均低于通过3D和组织学图像获得的均值/中值。对于组织学和显微CT,所有观察者的观察者内部ICC值最高。此外,组织学和CBCT的观察者内ICC值高于组织学和口腔内方法。对于所有观察者来说,一读和二读的观察者之间ICC值都很高。在Accuitomo和Iluma CBCT图像之间或口腔内膜和CCD图像之间,没有发现重复性差异。发现Micro-CT是离体测量咬合龋深度的最佳成像方法。此外,两个CBCT单元的表现均相似且优于口内模态。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号