Experienced physicians noted unexpectedly elevated concentrations of lactate dehydrogenase in some patient samples, but quality control specimens showed no bias. To evaluate this problem, we used a "latent reference individual extraction method", designed to obtain reference intervals from a laboratory database by excluding individuals who have abnormal results for basic analytes other than the analyte in question, in this case lactate dehydrogenase. The reference interval derived for the suspected year was 264-530 U/L, while that of the previous year was 248-495 U/L. The only change we found was the introduction of an order entry system, which requests precise sampling volumes rather than complete filling of vacuum tubes. The effect of vacuum persistence was tested using ten freshly drawn blood samples. Compared with complete filling, 1/5 filling resulted in average elevations of lactate dehydrogenase, aspartic aminotransferase, and potassium levels of 8.0%, 3.8%, and 3.4%, respectively (all p<0.01). Microhemolysis was confirmed using a urine stick method. The length of time before centrifugation determined the degree of hemolysis, while vacuum during centrifugation did not affect it. Microhemolysis is the probable cause of the suspected pseudo-elevation noted by the physicians. Data-mining methodology represents a valuable tool for monitoring long-term bias in laboratory results.
机译:经验丰富的医生指出,某些患者样本中的乳酸脱氢酶浓度出乎意料地升高,但质量控制样本未显示出偏倚。为了评估此问题,我们使用了一种“潜在参考个体提取方法”,该方法旨在通过排除对所讨论的分析物(在本例中为乳酸脱氢酶)以外的碱性分析物有异常结果的个人,从实验室数据库中获取参考间隔。怀疑年份的参考间隔为264-530 U / L,而上一年的参考间隔为248-495 U / L。我们发现的唯一变化是引入了订单输入系统,该系统要求精确的采样量而不是完全填充真空管。使用十个新鲜抽取的血液样本测试了真空持久性的影响。与完全填充相比,1/5填充导致乳酸脱氢酶,天冬氨酸转氨酶和钾水平的平均升高分别为8.0%,3.8%和3.4%(所有p <0.01)。使用尿棒法确认了微溶血。离心之前的时间长度决定了溶血的程度,而离心过程中的真空度对此没有影响。微溶血是医生指出的疑似假高的可能原因。数据挖掘方法学是监测实验室结果中长期偏见的宝贵工具。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号