首页> 外文期刊>Journal of clinical densitometry >In vivo precision of the GE Lunar iDXA for the assessment of lumbar spine, total hip, femoral neck, and total body bone mineral density in severely obese patients
【24h】

In vivo precision of the GE Lunar iDXA for the assessment of lumbar spine, total hip, femoral neck, and total body bone mineral density in severely obese patients

机译:GE Lunar iDXA的体内精确度,用于评估严重肥胖患者的腰椎,全髋,股骨颈和全身骨矿物质密度

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

No study has evaluated the precision of the GE Lunar iDXATM (GE Healthcare) in measuring bone mineral density (BMD) among severely obese patients. The purpose of the study was to evaluate the precision of the GE Lunar iDXATM for assessing BMD, including the lumbar spine L1-L4, L2-L4, the total hip, femoral neck, and total body in a severely obese population (body mass index [BMI] 40kg/m2). Sixty-four severely obese participants with a mean age of 46±11yr, BMI of 49±6kg/m2, and a mean body mass of 136.8±20.4kg took part in this investigation. Two consecutive iDXA scans (with repositioning) of the total body (total body BMD [TBBMD]), lumbar spine (L1-L4 and L2-L4), total hip (total hip BMD [THBMD]), and femoral neck (femoral neck BMD [FNBMD]) were conducted for each participant. The coefficient of variation (CV), the root mean square (RMS) averages of standard deviations of repeated measurements, the corresponding 95% least significant change, and intraclass correlations (ICCs) were calculated. In addition, analysis of bias and coefficients of repeatability were calculated. The results showed a high level of precision for total body (TBBMD), lumbar spine (L1-L4), and total hip (THBMD) with values of RMS: 0.013, 0.014, and 0.011g/cm2; CV: 0.97%, 1.05%, and 0.99%, respectively. Precision error for the femoral neck was 2.34% (RMS: 0.025g/cm2) but still represented high reproducibility. ICCs in all dual-energy X-ray absorptiometry measurements were 0.99 with FNBMD having the lowest at 0.98. Coefficients of repeatability for THBMD, FNBMD, L1-L4, L2-L4, and TBBMD were 0.0312, 0.0688, 0.0383, 0.0493, and 0.0312g/cm2, respectively. The Lunar iDXA demonstrated excellent precision for BMD measurements and is the first study to assess reproducibility of the GE Lunar iDXA with severely obese adults.
机译:尚无研究评估GE Lunar iDXATM(GE Healthcare)在严重肥胖患者中测量骨矿物质密度(BMD)的准确性。该研究的目的是评估GE Lunar iDXATM评估严重肥胖人群BMD的准确性,包括腰椎L1-L4,L2-L4,总髋部,股骨颈和全身(体重指数) [BMI]> 40kg / m2)。参加这项研究的平均年龄为46±11岁,BMI为49±6kg / m2,平均体重为136.8±20.4kg的64位严重肥胖参与者。两次连续iDXA扫描(重新定位):全身(全身BMD [TBBMD]),腰椎(L1-L4和L2-L4),全髋(全髋BMD [THBMD])和股骨颈(股骨颈)对每个参与者进行了BMD [FNBMD]。计算变异系数(CV),重复测量的标准偏差的均方根(RMS)平均值,相应的95%最低显着性变化以及类内相关性(ICC)。此外,还计算了偏差分析和重复性系数。结果显示,全身(TBBMD),腰椎(L1-L4)和全髋关节(THBMD)的精度较高,RMS值分别为0.013、0.014和0.011g / cm2; CV:分别为0.97%,1.05%和0.99%。股骨颈的精度误差为2.34%(RMS:0.025g / cm2),但仍具有很高的重现性。所有双能X射线吸收测定法中的ICC为0.99,而FNBMD最低,为0.98。 THBMD,FNBMD,L1-L4,L2-L4和TBBMD的重复性系数分别为0.0312、0.0688、0.0383、0.0493和0.0312g / cm2。 Lunar iDXA对BMD的测量显示出极好的精度,并且是第一个评估GE Lunar iDXA对严重肥胖成年人的可重复性的研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号