Hold-and-flush, a novel fraction collection method in semi-preparative subcritical and supercritical fluid chromatography

机译:保持冲洗,半制备亚临界和超临界流体色谱中的新型馏分收集方法

获取原文
获取原文并翻译 | 示例
           

摘要

In supercritical fluid chromatography, the partial phase separation that occurs in the dead volume between automatic backpressure regulator and collection vessels causes significant peak tailing and delayed arrival of compounds to the collectors. As a result, when two peaks are barely baseline separated, it becomes very difficult to correctly set fraction collection triggers, which in turn results in fractions being collected with lower purity and lower yield. The problem can be solved with a simple addition of a four-port, two-position switching valve between UV detector and automatic back pressure regulator. The valve acts as a timed gate to release each peak out of the back pressure regulator and into the collection vessel, while at the same time it holds later peaks inside the UV detector and the column in a closed loop. (C) 2004 Elsevier B.V. All rights reserved.
机译:在超临界流体色谱中,在自动背压调节器和收集容器之间的死区中发生的部分相分离会导致明显的峰拖尾和化合物到达收集器的延迟到达。结果,当两个峰几乎没有基线分离时,很难正确设置馏分收集触发信号,从而导致以较低的纯度和较低的收率收集馏分。只需在紫外线检测器和自动背压调节器之间增加一个四通二位切换阀即可解决该问题。该阀用作定时闸门,以将每个峰从背压调节器中释放出来并进入收集容器,同时,它将以后的峰保持在UV检测器和色谱柱内为闭环。 (C)2004 Elsevier B.V.保留所有权利。