Evaluation of peripheral pulmonary arteries at 80 kV and at 140 kV: dual-energy computed tomography assessment in 51 patients.

机译:在80 kV和140 kV下评估周围肺动脉:51位患者的双能计算机断层扫描评估。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To compare peripheral pulmonary artery image quality at 80 kVp and 140 kVp in the same patients. MATERIALS AND METHODS: Image quality of third-, fourth- and fifth-order arteries was assessed at 80 kV and 140 kV on 1-mm-thick transverse scans, generated from dual-source computed tomography (CT) acquisitions. RESULTS: The mean level of enhancement was significantly higher at 80 kV compared with 140 kV for the third-, fourth-, and fifth-order arteries (P < 0.0001). Despite a higher noise level at 80 kV (P < 0.0001), the signal-to-noise ratio and contrast-to-noise ratio were significantly higher at 80 kV than at 140 kV at the level of third-, fourth-, and fifth-order arteries (P < 0.0001). The mean vascular attenuation, mean signal-to-noise ratio and contrast-to-noise ratio for peripheral arteries were significantly superior at 80 kV in the 3 body mass index categories (P < 0.005). CONCLUSIONS: Eighty-kilovolt protocols significantly improve the image quality of peripheral pulmonary arteries on CT angiograms of the chest.
机译:目的:比较同一患者在80 kVp和140 kVp时周围肺动脉的图像质量。材料与方法:通过双源计算机断层扫描(CT)采集,在1毫米厚的横向扫描中评估了80 kV和140 kV时三阶,四阶和五阶动脉的图像质量。结果:80 kV的平均增强水平明显高于三,四和五阶动脉的140 kV(P <0.0001)。尽管在80 kV时具有较高的噪声水平(P <0.0001),但在第三,第四和第五级时,在80 kV时的信噪比和对比度-噪声比明显高于在140 kV时的信噪比和对比度有序动脉(P <0.0001)。在3种体重指数类别中,在80 kV时,外周动脉的平均血管衰减,平均信噪比和对比噪声比明显更好(P <0.005)。结论:80千伏特协议可显着改善胸部CT血管造影上周围肺动脉的图像质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号