Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems.

机译:通过Galilei和Orbscan II分析系统获得的角膜厚度和曲率读数的比较和可重复性。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To compare central corneal thickness (CCT) and keratometry readings using the Galilei dual Scheimpflug analyzer, the Orbscan II anterior segment analysis system, and the Sonogage ultrasound (US) pachymeter. SETTING: Lucerne Eye Clinic, Cantonal Hospital, Lucerne, Switzerland. METHODS: In a prospective single-center study, 85 eyes of 45 healthy volunteers were examined with the Orbscan II and Galilei systems in random order followed by Sonogage US pachymetry. The CCT and keratometry measurements were compared using a paired t test. To assess the intraobserver and interobserver reproducibility of the 2 keratometers, 9 volunteers were examined 3 times by 4 independent observers and the intraclass correlation coefficient (ICC) was calculated. RESULTS: The mean CCT was 551.7 microm+/-36.6 (SD) with the Galilei, 554.8+/-45.1 microm with the Orbscan II, and 558.5+/-38.4 microm with the Sonogage. The CCT readings of the Galilei and Orbscan II did not differ significantly (P= .12). The mean keratometry readings with the Galilei and Orbscan II were similar, although both the steep (Ks) and flat (Kf) axes tended to be flatter with the Galilei system. The ICCs for CCT, Kf, and Ks were high with both keratometers (0.98, 0.97, 0.84 for Galilei and 0.97, 0.96, 0.95 for Orbscan II, respectively), indicating that variation in measurements was mainly due to true subject-to-subject variation rather than observer error. CONCLUSION: Keratometry and pachymetry readings with the Galilei and Orbscan II systems showed good concordance and high reproducibility, which would allow the examinations to be delegated to nonmedical personnel.
机译:目的:使用伽利略双Scheimpflug分析仪,Orbscan II前节分析系统和Sonogage超声(美国)测厚仪比较中央角膜厚度(CCT)和角膜曲率法读数。地点:瑞士卢塞恩州立医院卢塞恩眼科诊所。方法:在一项前瞻性单中心研究中,对45位健康志愿者的85只眼睛进行了随机检查,分别采用Orbscan II和Galilei系统,然后进行了Sonogage US测厚仪。使用配对t检验比较CCT和角膜曲率测量。为了评估2个角膜曲率计的观察者内部和观察者之间的可重复性,由4位独立观察员对9名志愿者进行了3次检查,并计算了组内相关系数(ICC)。结果:Galilei的平均CCT为551.7微米+/- 36.6微米(SD),Orbscan II的平均CCT为554.8 +/- 45.1微米,Sonogage的平均CCT为558.5 +/- 38.4微米。 Galilei和Orbscan II的CCT读数无显着差异(P = .12)。 Galilei和Orbscan II的平均角膜曲率读数相似,尽管陡峭(Ks)和平坦(Kf)轴在Galilei系统中趋于平坦。两种角膜曲率计的CCT,Kf和Ks的ICC均较高(伽利略仪的分别为0.98、0.97、0.84,Orbscan II的分别为0.97、0.96、0.95),这表明测量值的变化主要是由于真实受试者之间的差异变化而不是观察者错误。结论:Galilei和Orbscan II系统的角膜测定法和测厚法读数显示出良好的一致性和高可重复性,这将使检查委托给非医务人员进行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号