Corneal thickness measurements in normal and keratoconic eyes: Pentacam comprehensive eye scanner versus noncontact specular microscopy and ultrasound pachymetry.

机译:正常和圆锥角膜眼睛的角膜厚度测量:Pentacam全面眼扫描仪与非接触式镜检显微镜和超声测厚仪。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To compare central corneal thickness (CCT) measurements taken with the Pentacam comprehensive eye scanner (CES), noncontact specular microscopy (SM), and ultrasound pachymetry (UP) in normal and keratoconic corneas. SETTING: Department of Ophthalmology, Ankara University School of Medicine, Ankara, Turkey. METHODS: In a prospective study, 3 CCT measurements were taken with the Pentacam CES, SM, and UP in that sequence from 1 eye of 45 consecutive patients with myopia (group A) and 62 consecutive patients with keratoconus (group B). Eyes with keratoconus were further divided into 3 subgroups, mild, moderate, and severe, according to the mean keratometry readings. RESULTS: Pentacam CES (r = 0.994) and UP (r = 0.993) demonstrated very high and comparable reproducibility in group A. In group B, Pentacam CES displayed better reproducibility (r = 0.988) than UP (r = 0.969) and SM (r = 0.901). The mean CCT measurements of Pentacam CES and UP were not significantly different in group A (P = .37) and in eyes with mild keratoconus (P = .29), whereas significant differences between all instrument pairs were evident in group B and in moderate and severely keratoconic eyes (P<.05). Noncontact SM measurements were significantly smaller than those of Pentacam CES and UP in all groups (P<.05) when a measurement could be obtained. There were significant linear correlations between CCT measurements of Pentacam CES, UP, and noncontact SM in groups A, B, and mildly keratoconic eyes (P<.001). CONCLUSIONS: Results suggest that whereas Pentacam CES and UP may be used interchangeably in normal eyes in the clinical setting for the measurement of CCT, one should be cautious interpreting corneal thickness data from Pentacam CES, UP, and particularly SM in eyes with keratoconus. Whereas, in normal and mildly keratoconic eyes, Pentacam CES and UP demonstrated very high and comparable reproducibility, in moderately keratoconic eyes, Pentacam CES readings showed better reproducibility than UP.
机译:目的:比较在正常角膜和圆锥角膜中使用Pentacam综合眼部扫描仪(CES),非接触式镜检显微镜(SM)和超声测厚仪(UP)进行的中央角膜厚度(CCT)测量。地点:土耳其安卡拉安卡拉大学医学院眼科。方法:在一项前瞻性研究中,使用Pentacam CES,SM和UP依次进行了3次CCT测量,分别来自45位连续的近视患者(A组)和62位连续的圆锥角膜患者(B组)的一只眼睛。根据平均角膜测量法的读数,圆锥角膜的眼睛进一步分为轻度,中度和重度三个亚组。结果:Pentacam CES(r = 0.994)和UP(r = 0.993)在A组中显示出非常高且可比的重现性。在B组中,Pentacam CES显示出比UP(r = 0.969)和SM( r = 0.901)。 Pentacam CES和UP的平均CCT测量值在A组(P = .37)和轻度圆锥角膜(P = .29)眼中无显着差异,而B组和中度在所有器械对之间均存在显着差异以及严重圆锥角膜的眼睛(P <.05)。当可以获得测量值时,所有组的非接触式SM测量值均显着小于Pentacam CES和UP(P <.05)。在A组,B组和轻度圆锥角膜眼中,Pentacam CES,UP和非接触式SM的CCT测量值之间存在显着的线性相关性(P <.001)。结论:结果表明,尽管Pentacam CES和UP在临床环境中可在正常眼睛中互换使用以测量CCT,但应谨慎解释来自Pentacam CES,UP的角膜厚度数据,尤其是圆锥角膜眼睛的SM。而在正常和轻度圆锥角膜眼中,Pentacam CES和UP表现出非常高且可比的重现性;在中度圆锥角膜眼中,Pentacam CES读数显示出比UP更好的重现性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号