PURPOSE: To determine the difference in pupil size measured with the Colvard pupillometer in mesopic and scotopic luminance and with the Rosenbaum pupil card in mesopic luminance between 2 examiners. SETTING: Michel Pop Clinics, Montreal, Quebec, Canada. METHODS: Two examiners used the Colvard pupillometer and the Rosenbaum card to measure pupil size in 58 eyes. The Colvard pupillometer was used in mesopic and scotopic light conditions. The Rosenbaum card was used in mesopic luminance only. Pupil size was evaluated with a 1.0 mm interval scale at the nearest half millimeter. RESULTS: For the 3 sets of data, the limits of agreement and coefficient of interrater repeatability were calculated and a 2 x 2 factorial analysis of variance was performed. Because of interexaminer bias, measurements done in mesopic luminance with the Rosenbaum card were not statistically different from those with the Colvard pupillometer in scotopic luminance, although interrater repeatability of the Colvard pupillometer (0.8 mm) was superior to that of the Rosenbaum card (1.3 mm). CONCLUSIONS: Examiner bias was the greatest statistical bias in all sets of measures. Surgeons may want to opt for a "safe" limit of pupil size (ie, 0.5 to 0.8 mm greater than the measured size) when calculating optical zones in refractive surgery. Future devices for pupil measurement should be based on automatic adjustment sizing.
机译:目的:确定2名检查者之间用Colvard瞳孔计测量的中视和暗视亮度以及用Rosenbaum瞳孔卡测量的中视亮度之间的差异。地点:加拿大魁北克蒙特利尔的米歇尔·波普诊所。方法:两名检查员使用Colvard瞳孔计和Rosenbaum卡测量58眼的瞳孔大小。 Colvard瞳孔计用于中视和暗视照明条件。 Rosenbaum卡仅用于介观亮度。用1.0mm间隔刻度在最接近的半毫米处评估瞳孔大小。结果:对于这三组数据,计算出一致性的限制和人际可重复性系数,并进行了2 x 2的方差分析。由于考官之间的偏见,尽管高氏瞳孔计的可重复性(0.8毫米)优于罗森鲍姆卡(1.3毫米)的可重复性,但使用Rosenbaum卡进行介观亮度的测量与使用Colvard瞳孔计进行的暗视亮度测量在统计学上没有差异。 )。结论:在所有指标中,考官偏倚是最大的统计偏倚。在屈光手术中计算光学区域时,外科医生可能希望选择瞳孔尺寸的“安全”限制(即,比测量尺寸大0.5至0.8毫米)。将来用于瞳孔测量的设备应基于自动调整大小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号