Combined occlusion-triggered longitudinal and torsional phacoemulsification during coaxial microincision cataract surgery: effect on 30-degree mini-flared tip behavior.

机译:同轴微切口白内障手术期间联合咬合触发的纵向和扭转超声乳化术:对30度微喇叭形尖端行为的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To study the impact of a new power-delivery option using occlusion-triggered bursts of longitudinal phaco energy in addition to torsional tip movement in the performance of the 30-degree mini-flared tip to minimize tip clogging with dense nuclei. SETTING: Prince Charles Eye Unit, King Edward VII Hospital, Windsor, United Kingdom. DESIGN: Comparative case series. METHOD: The cumulative dissipated energy (CDE) and time in foot pedal position 3 (FP3) to phacoemulsify each half of the nucleus with and without use of the new power-delivery option (Intelligent Phaco) were compared. RESULTS: Fifty consecutive cataract operations were reviewed. The mean CDE was 1.79 with the new power-delivery option and 2.03 without the option (P=.08). For grades I, II, III, and IV nuclei, the corresponding values were 0.41 and 0.51 (P=.16), 1.43 and 1.23 (P=.31), 2.32 and 2.92 (P=.12), 3.82 and 5.03 (P=.005). The mean FP 3 time was 5.40 seconds with the new power-delivery option and 6.35 seconds without it (P=.03). For grades I, II, III, and IV nuclei, the corresponding values were 1.53 seconds and 1.98 seconds (P=.13), 4.47 seconds and 4.26 seconds (P=.69), 6.51 seconds and 8.77 seconds (P=.05), 11.81 seconds and 16.49 seconds (P=.008). CONCLUSIONS: The new power-delivery option appeared to reduce the CDE and FP 3 time when a 30-degree mini-flared tip was used. The benefit accrued by preventing clogging of the needle was significant only in cases with a denser nucleus.
机译:用途:研究使用新的动力传递选项的影响,该特性使用了扭转相位运动的纵向波可能量的咬合触发脉冲串,并在30度微型喇叭形尖端的性能中发挥了最大作用,以最大程度地减少密集核的尖端堵塞。地点:英国温莎爱德华七世国王医院查尔斯王子眼科。设计:比较案例系列。方法:比较了使用和不使用新的功率传递选项(Intelligent Phaco)时脚踏板位置3(FP3)的超声耗散能量(CDE)和时间,以超声乳化每半个核。结果:连续五十次白内障手术进行了审查。使用新的功率传输选项的平均CDE为1.79,不使用此选项的平均CDE为2.03(P = .08)。对于I,II,III和IV级核,相应值分别为0.41和0.51(P = .16),1.43和1.23(P = .31),2.32和2.92(P = .12),3.82和5.03( P = .005)。 FP 3的平均时间在使用新的电源传输选项时为5.40秒,而在没有此选项的情况下为6.35秒(P = .03)。对于I,II,III和IV级核,相应的值为1.53秒和1.98秒(P = .13),4.47秒和4.26秒(P = .69),6.51秒和8.77秒(P = .05) ),11.81秒和16.49秒(P = .008)。结论:当使用30度微型喇叭形尖端时,新的功率传送选项似乎减少了CDE和FP 3的时间。仅在细胞核较密的情况下,预防针头堵塞所带来的好处才是重要的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号