首页> 外文期刊>Journal of cataract and refractive surgery >Rotational and centration stability of an aspheric intraocular lens with a simulated toric design.
【24h】

Rotational and centration stability of an aspheric intraocular lens with a simulated toric design.

机译:具有模拟复曲面设计的非球面人工晶状体的旋转和定心稳定性。

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: To assess the stability of the Akreos AO intraocular lens (IOL) platform with a simulated toric design using objective image analysis. SETTING: Six hospital eye clinics across Europe. METHODS: After implantation in 1 eye of patients, IOLs with orientation marks were imaged at 1 to 2 days, 7 to 14 days, 30 to 60 days, and 120 to 180 days. The axis of rotation and IOL centration were objectively assessed using validated image analysis. RESULTS: The study enrolled 107 patients with a mean age of 69.9 years +/- 7.7 (SD). The image quality was sufficient for IOL rotation analysis in 91% of eyes. The mean rotation between the first day postoperatively and 120 to 180 days was 1.93 +/- 2.33 degrees, with 96% of IOLs rotating fewer than 5 degrees and 99% rotating fewer than 10 degrees. There was no significant rotation between visits and no clear bias in the direction of rotation. In 71% of eyes, the dilation and image quality was sufficient for image analysis of centration. The mean change in centration between 1 day and 120 to 180 days was 0.21 +/- 0.11 mm, with all IOLs decentering less than 0.5 mm. There was no significant decentration between visits and no clear bias in the direction of the decentration. CONCLUSION: Objective analysis of digital retroillumination images taken at different postoperative periods shows the aspheric IOL platform was stable in the eye and is therefore suitable for the application of a toric surface to correct corneal astigmatism.
机译:目的:使用客观图像分析的模拟复曲面设计评估Akreos AO人工晶状体(IOL)平台的稳定性。地点:全欧洲六家医院眼科诊所。方法:在1只患者的眼睛中植入后,在1至2天,7至14天,30至60天和120至180天对带有方向标记的IOL成像。使用经过验证的图像分析客观地评估旋转轴和IOL中心。结果:该研究招募了107名平均年龄为69.9岁+/- 7.7(SD)的患者。 91%的眼睛的图像质量足以进行IOL旋转分析。术后第一天至120至180天的平均旋转度为1.93 +/- 2.33度,其中96%的IOL旋转度小于5度,而99%的IOL旋转度小于10度。访视之间没有明显的旋转,旋转方向也没有明显的偏差。在71%的眼睛中,扩张和图像质量足以对定点进行图像分析。 1天到120至180天之间的平均中心变化为0.21 +/- 0.11毫米,所有IOL的偏心距均小于0.5毫米。探访之间没有明显的分权,分权的方向也没有明显的偏见。结论:客观分析术后不同时期拍摄的数字后照图像显示非球面人工晶状体平台在眼内稳定,因此适合使用复曲面来矫正角膜散光。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号