Performance of infrared ear and forehead thermometers: A comparative study in 205 febrile and afebrile children

机译:红外耳温计的性能:205名高热和低热儿童的比较研究

获取原文
获取原文并翻译 | 示例

摘要

Aims and objectives: This study compared readings from two professional-grade, commercially available infrared (IR) thermometers, the ThermoScan? PRO 4000 prewarmed tip ear thermometer and the Temporal Scanner? TAT-5000 temporal artery thermometer. Background: The repeatability and precision of readings from IR thermometers for professional use were questioned in the past, but in recent years, these types of thermometers have been technologically improved, so their ability to replicate standard temperature readings reliably should be re-examined. Design: Febrile and afebrile children were recruited from the emergency department, overflow treatment areas and the paediatric intensive care unit of a large hospital in Argentina. Each child had a randomised sequence of seven temperature readings, including three from the ear, three from the forehead or behind the ear and one reference oral or rectal reading. Methods: Temperature readings were taken with the ThermoScan PRO 4000, the Temporal Scanner TAT-5000 and the monitor mode of SureTemp? Plus, a widely used professional-grade contact thermometer, for reference. Results: Of 205 children, 46% were febrile, per reference thermometer readings. While mean ThermoScan PRO 4000 febrile measurements did not differ significantly from reference, mean Temporal Scanner TAT-5000 febrile measurements were significantly lower (by a mean of 0·42 °C) than the reference. Overall bias of the ThermoScan PRO 4000 was significantly lower than that of the Temporal Scanner TAT-5000; repeatability was 1·5 times higher, and overall false-negative rate was about a third that of Temporal Scanner TAT-5000, when compared to the reference. Conclusions: This study indicates that the ThermoScan PRO 4000 provides measurements closer to those of a professional-grade contact thermometer when compared to the Temporal Scanner TAT-5000. Relevance to clinical practice: The odds of misclassifying a febrile child as non-febrile were about threefold higher with the Temporal Scanner TAT-5000 temporal artery thermometer than with the ThermoScan? PRO 4000 ear thermometer.
机译:目的和目的:本研究比较了两个专业级的,可商购的红外(IR)温度计ThermoScan?的读数。 PRO 4000预热的尖端耳温计和Temporal Scanner? TAT-5000颞动脉温度计。背景:过去,对用于专业用途的红外温度计的读数的可重复性和准确性提出了质疑,但是近年来,这类温度计已在技术上得到了改进,因此应重新检查其可靠地复制标准温度读数的能力。设计:从阿根廷一家大型医院的急诊科,溢流治疗区和儿科重症监护室招募发热和无发热的儿童。每个孩子都有七个温度读数的随机序列,其中三个来自耳朵,三个来自额头或耳朵后面,以及一个参考口头或直肠读数。方法:使用ThermoScan PRO 4000,Temporal Scanner TAT-5000和SureTemp?的监控模式获取温度读数。另外,广泛使用的专业级接触式温度计可供参考。结果:根据参考体温表读数,在205名儿童中,有46%为高热。尽管平均ThermoScan PRO 4000的发热测量值与参考值无显着差异,但平均Temporal Scanner TAT-5000的发热测量值比参考值要低得多(平均值为0·42°C)。 ThermoScan PRO 4000的总体偏差明显低于Temporal Scanner TAT-5000;与参考相比,可重复性高1·5倍,总体假阴性率约为Temporal Scanner TAT-5000的三分之一。结论:这项研究表明,与Temporal Scanner TAT-5000相比,ThermoScan PRO 4000提供的测量值更接近于专业级接触式温度计。与临床实践的相关性:使用Temporal Scanner TAT-5000颞动脉温度计将发热儿童误分类为非发热的几率比ThermoScan高三倍。 PRO 4000耳温计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号