首页> 外文期刊>Clinical biomechanics >Importance of polyethylene thickness in total shoulder arthroplasty: A finite element analysis
【24h】

Importance of polyethylene thickness in total shoulder arthroplasty: A finite element analysis

机译:聚乙烯厚度在全肩关节置换术中的重要性:有限元分析

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Articular surfaces reconstruction is essential in total shoulder arthroplasty. Because of the limited glenoid bone support, thin glenoid component could improve anatomical reconstruction, but adverse mechanical effects might appear. Methods: With a numerical musculoskeletal shoulder model, we analysed and compared three values of thickness of a typical all-polyethylene glenoid component: 2, 4 (reference) and 6 mm. A loaded movement of abduction in the scapular plane was simulated. We evaluated the humeral head translation, the muscle moment arms, the joint force, the articular contact pattern, and the polyethylene and cement stress. Findings Decreasing polyethylene thickness from 6 to 2 mm slightly increased humeral head translation and muscle moment arms. This induced a small decreased of the joint reaction force, but important increase of stress within the polyethylene and the cement mantel. Interpretation The reference thickness of 4 mm seems a good compromise to avoid stress concentration and joint stuffing.
机译:背景:关节面重建在全肩关节置换术中至关重要。由于关节盂骨支撑有限,关节盂变薄可以改善解剖结构,但可能会出现不利的机械作用。方法:使用数值型的骨骼骨骼肩关节模型,我们分析并比较了典型的全聚乙烯类关节盂组件的三个厚度值:2、4(参考)和6mm。模拟了肩duction平面内外展的负重运动。我们评估了肱骨头的平移,臂的肌肉力矩,关节力,关节接触方式以及聚乙烯和水泥的应力。研究结果聚乙烯厚度从6毫米减少到2毫米,略微增加了肱骨头平移和肌肉力矩臂。这引起了联合反作用力的小幅下降,但聚乙烯和水泥壁炉架内的应力却显着增加。解释4 mm的参考厚度似乎是一个很好的折衷方案,可以避免应力集中和接头堵塞。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号