Objectives: We assessed the analytical performance of the TSH and FT4 assays on ADVIA Centaur in a multicenter national evaluation. Design and methods: A precision study and a method comparison were performed. Reference values stated by the manufacturer were checked from 379 normal subjects. Results: For TSH and FT4, the intra-assay CVs were below 2.3 and 5.2%, respectively, and the inter-assay CVs below 4.4% and 7.2%, respectively. Therefore, the precision and reproducibility were acceptable. Bland-Altman bias plots revealed good correlation and agreement with Cobas assays. TSH and FT4 data yielded reference ranges of 0.64-3.24. mIU/L and 10.5-18.9. pmol/L, respectively. Conclusion: These assays demonstrate reliable characteristics. The reference ranges obtained can be used for interpretation of thyroid function.
机译:目标:我们在一项多中心国家评估中评估了TDVI和FT4分析在ADVIA Centaur上的分析性能。设计和方法:进行了精密度研究和方法比较。从379名正常受试者中检查了制造商规定的参考值。结果:对于TSH和FT4,批内CV分别低于2.3%和5.2%,批间CV分别低于4.4%和7.2%。因此,精密度和再现性是可以接受的。 Bland-Altman偏倚图显示与Cobas分析具有良好的相关性和一致性。 TSH和FT4数据得出的参考范围为0.64-3.24。 mIU / L和10.5-18.9。 pmol / L。结论:这些测定证明了可靠的特征。获得的参考范围可用于解释甲状腺功能。