首页> 外文期刊>Journal of cardiovascular medicine >Conductor externalization of the RIATA implantable cardioverter defibrillator lead: A challenging lead extraction requiring laser-powered sheath
【24h】

Conductor externalization of the RIATA implantable cardioverter defibrillator lead: A challenging lead extraction requiring laser-powered sheath

机译:RIATA植入式心脏复律除颤器引线的导体外化:极富挑战性的引线提取,需要激光供电的护套

获取原文
获取原文并翻译 | 示例
       

摘要

We report the case of a 27-year-old man who received a single chamber implantable cardioverter defibrillator with a St Jude Medical Riata double coil passive lead for secondary prevention of sudden cardiac death due to idiopathic ventricular fibrillation. Subsequently, he was readmitted for lead extraction because of device infection. Fluoroscopy showed externalization of the conductors proximally to the distal coil in the absence of any signs of electrical lead failure. Successful lead extraction was achieved using a laser sheath without any complications after unsuccessful mechanical sheath because of the tenacious fibrotic adherences proximal to the distal coil and extrusion of the conductors that hindered the advancement of the sheath over the lead. Our report highlights that Riata ICD leads are prone to externalization of the conductor wires, which may render an indicated lead extraction procedure of these leads more challenging. A laser powered sheath may facilitate the procedure, especially in the presence of extensive adhesions and fibrosis.
机译:我们报道了一名27岁男子的案例,该男子接受了单腔植入式心脏复律除颤器和St Jude Medical Riata双线圈被动导线,用于继发性预防因特发性心室颤动而导致的心脏猝死。随后,由于设备感染,他被重新接纳了铅提取。荧光检查显示,在没有任何电引线故障迹象的情况下,导线在远侧线圈的近端外部化。在不成功的机械护套之后,使用激光护套可以成功地拔出铅,而不会产生任何并发症,这是因为靠近远端线圈的顽强的纤维化粘附以及导体的挤出阻碍了护套在导线上的前进。我们的报告强调说,Riata ICD引线易于导线外部化,这可能会使这些引线的指示引线抽取过程更具挑战性。激光驱动的护套可以促进手术,特别是在存在广泛的粘连和纤维化的情况下。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号