Characterization, quantification, and isolation of aluminum oxide particles on grit blasted titanium aluminum alloy hip implants.

机译:喷砂处理的钛铝合金髋关节植入物中氧化铝颗粒的表征,定量和分离。

获取原文
获取原文并翻译 | 示例

摘要

BACKGROUND: This study was undertaken to verify whether or not the microstructure of aluminum alloy implants interferes with the characterization and quantification of aluminum inclusions on their surfaces, resulting from grit blasting. METHODS: Four factory-fresh prostheses were investigated by scanning electron microscopy and X-ray microanalysis. Specimens were cut out of the stems and the cross-sections analyzed. The specimens were etched in hot 25% hydrochloric acid. The hydrochloric acid was subsequently filtered with a 0.2-microm-pore filter. The filters were scanned using electron microscopy and X-ray microanalysis. RESULTS: Aluminum oxide particles were found on all investigated stems; the diameter of the particles ranged from 4 to 100 microm. One hundred fifty-four particles were counted per mm(2). No particles were seen on the cross-sections of the implants. Scanning electron microscopy of the Millipore filters revealed aluminum oxide particles. CONCLUSION: Remnants of grit blasting were found only on the surfaces; none were observed on cross-sections. We conclude that the microstructure of titanium aluminum alloy does not interfere with the identification and quantification of particles. Particles were identified on the filters by electron microscopy and X-ray microanalysis. Aluminum oxide on the surface of grit-blasted titanium aluminum alloy implants is, in fact, a residue of grit blasting.
机译:背景:这项研究旨在验证铝合金植入物的微观结构是否会干扰喷砂处理导致表面上铝夹杂物的表征和定量。方法:通过扫描电子显微镜和X射线显微分析法研究了四种出厂的假体。从茎上切下标本并分析横截面。样品在热的25%盐酸中蚀刻。随后用0.2微米孔过滤器过滤盐酸。使用电子显微镜和X射线微分析仪扫描滤光片。结果:在所有研究的茎上均发现了氧化铝颗粒。颗粒的直径为4至100微米。每毫米计数154个粒子(2)。在植入物的横截面上没有看到颗粒。密理博过滤器的扫描电子显微镜显示出氧化铝颗粒。结论:仅在表面发现了喷砂残留物。在横截面上没有观察到。我们得出的结论是,钛铝合金的微观结构不会干扰颗粒的鉴定和定量。通过电子显微镜和X射线显微分析在过滤器上鉴定出颗粒。实际上,喷砂处理的钛铝合金植入物表面的氧化铝实际上是喷砂处理的残留物。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号