Electromagnetic interference of implantable cardiac devices from a shoulder massage machine.

机译:肩部按摩机对植入式心脏设备的电磁干扰。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter-defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28?cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a "don't use" warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.
机译:肩部按摩器有两个垫板,由电磁线圈驱动,可对肩膀进行每次敲打按摩。由于电磁线圈中的交流电会产生随时间变化的磁场,因此人们担心此类机器可能会在植入式心脏设备中产生电磁干扰(EMI)。这项研究的目的是研究一台此类肩部按摩机在植入式心脏设备上的潜在EMI。我们测量了按摩机周围磁场强度的分布曲线。此外,我们使用标准化的Irnich人体模型对植入式心脏起搏器进行了抑制试验和异步试验。我们在按摩机运行时使用起搏器编程器检查了植入式心脏复律除颤器(ICD)上的事件。磁场分布曲线在一个电磁线圈中显示出212(A / m)的峰值强度。按摩机与植入式心脏起搏器之间的最大干扰距离为28?cm。当按摩机靠近ICD的电极并触及Irnich人体模型时,就会诱发心室纤颤。有必要在按摩器的包装盒或外部上或在用户手册中提供“请勿使用”警告,并提醒植入起搏器的患者注意按摩器的危险和正确使用方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号