摘要

Insertion of cannulae into vessels may apply non-physiological load and stress on blood cells, such that adenosine diphosphate may increase and result in hemolysis. Authors used the computational method to simulate the blood flow inside of the cannula. We limited the research to within the drainage cannulae. Nine different cannulae categorized by the number of side holes of 4, 12, and 20, and also categorized by the array type as staggered array, in-line array, and alternative in-line array were studied and compared to the cannulae with no side holes by using computational fluid dynamics. We evaluated the flow rate, the wall shear stress, and the shear rate, and compared them with one another to estimate the effect of the side holes. The flow rate is not proportional to the number of the side holes. However, larger number of side holes can reduce the mean shear rate.
机译:将套管插入血管可能会对血细胞施加非生理负荷和压力,以致二磷酸腺苷可能增加并导致溶血。作者使用计算方法来模拟套管内的血流。我们的研究仅限于引流套管内。研究了9个按侧孔数4、12和20划分的不同插管,还按阵列类型分为交错阵列,直列阵列和替代直列阵列,并将其与无侧插管进行了比较通过使用计算流体动力学来钻孔。我们评估了流量,壁面剪切应力和剪切速率,并将它们相互比较以评估侧孔的影响。流量与侧孔的数量不成比例。但是,较大数量的侧孔会降低平均剪切速率。