Reproducibility of the pressure biofeedback unit in measuring transversus abdominis muscle activity in patients with chronic nonspecific low back pain

机译:压力生物反馈装置在慢性非特异性下腰痛患者中测量腹横肌活动的可重复性

获取原文
获取原文并翻译 | 示例
           

摘要

The Pressure Biofeedback Unit (PBU) is often used by clinicians and researchers to indirectly evaluate transversus abdominis (TrA) muscle activity. The purpose of this study was to evaluate the inter and intra-examiner reproducibility of the PBU in measuring TrA muscle activity in fifty patients with chronic nonspecific low back pain. This study was performed using a test-retest design with a seven day interval. An Intraclass Correlation Coefficient (ICC 2,1) of 0.74 (95% CI 0.54 to 0.85) and 0.76 (95% CI 0.58 to 0.86) was observed for the intra and inter-examiner reproducibility, respectively. The intra-examiner agreement (Limits of Agreement - LOA = 2.1 to -1.8 mmHg) and the inter-examiner agreement (LOA = 2.0 to -1.9 mmHg) were within the limits of agreement on 95% of occasions. The reproducibility of PBU in measuring TrA muscle activity in patients with chronic nonspecific low back pain ranged from satisfactory to excellent.
机译:临床医生和研究人员经常使用压力生物反馈装置(PBU)间接评估腹横肌(TrA)的肌肉活动。这项研究的目的是评估在测量50例慢性非特异性下背痛患者中TrA肌肉活动时,PBU在检查者内部和内部的可重复性。这项研究使用间隔7天的重新测试设计进行。组内和组间可重复性的内部相关系数(ICC 2,1)分别为0.74(95%CI 0.54至0.85)和0.76(95%CI 0.58至0.86)。检验者内部协议(协议的限制-LOA = 2.1至-1.8 mmHg)和检验者内部协议(LOA = 2.0至-1.9 mmHg)在95%的情况下均在协议范围内。在慢性非特异性下背痛患者中,PBU在测量TrA肌肉活动中的重现性从令人满意到出色。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号