A microcolumn of alumina modified with sodium dodecyl sulfate (SDS) and 1-(2-pyridylazo)-2-naphthol (PAN) was prepared for the preconcentration of trace nickel from water samples for a flame atomic absorption spectrometry (FAAS) determination. Under optimized conditions (pH = 4.0; flow rate, 5 mL min~(–1)) nickel (II) was retained on the column. The nickel collected on the column was eluted with 5 mL of 0.5 M nitric acid. Recovery was greater than 96.7%. A concentration factor of 300 can be achieved by passing 1500 mL of sample through the microcolumn. The relative standard deviation (ten replicate analyses) at the 40 ng mL~(–1) level for nickel was 2.4%, and the corresponding limit of detection (based on 3) was 0.06 ng mL~(–1). The method was applied to the determination of Ni in waste and mineral waters.
机译:制备了用十二烷基硫酸钠(SDS)和1-(2-吡啶基偶氮)-2-萘酚(PAN)改性的氧化铝微柱,用于从水样中富集痕量镍,用于火焰原子吸收光谱法(FAAS)测定。在优化的条件下(pH = 4.0;流速,5 mL min〜(–1)),镍(II)保留在色谱柱上。用5 mL的0.5 M硝酸洗脱柱上收集的镍。回收率大于96.7%。通过使1500 mL样品通过微柱可达到300的浓缩系数。镍在40 ng mL〜(–1)浓度下的相对标准偏差(十次重复分析)为2.4%,相应的检出限(基于3)为0.06 ng mL〜(–1)。该方法用于废水和矿泉水中镍的测定。