Performance evaluation of three on-site adulterant detection devices for urine specimens.

机译:三种尿液现场掺假检测装置的性能评估。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The performance of three on-site adulterant detection devices that assess the integrity of urine specimens collected for drug-of-abuse testing was evaluated: the Intect 7, MASK Ultra Screen, and Adultacheck 4. Intect 7 simultaneously tests creatinine, nitrite, glutaraldehyde, pH, specific gravity, and the presence of bleach and pyridinium chlorochromate (PCC). Mask Ultra Screen tests creatinine, nitrite, pH, specific gravity, and oxidants, and Adultacheck 4 tests creatinine, nitrite, glutaraldehyde, and pH. Urine specimens were prepared with the Substance Abuse and Mental Health Administration regulated analytes at 50% above the cut-off concentrations. Stealth, Urine Luck, Instant Clean ADD-IT-ive, and KLEAR were added individually to the drug-added urine specimens so that their concentrations reflected the "optimum" usage reported in their package inserts and 25% above and below that optimum. Stealth is reported to be peroxidase; Urine Luck is believed to be PCC; Instant Clean ADD-it-ive reportedly contains glutaraldehyde, and Klear is a nitrite. The following diluents/adulterants were added at 25%, 33%, and 50% of the volume of drug-added urine: distilled water, bleach, ammonia, and vinegar. Of the devices tested, Intect 7 proved to be the most sensitive, and it correctly indicated the presence of adulterant or diluent in all samples tested. In order to do so, all indication pads had to be assessed in concert. Adultacheck 4 specifically assesses four characteristics of urine integrity and is therefore very limited in detecting the use of several popular adulterants that are commercially available. Although it correctly assessed the four characteristics, it did not detect the use of Stealth, Urine Luck, or Instant Clean ADD-it-ive. Mask Ultra Screen can potentially detect a broader range of adulterants than Adultacheck 4. However, in practice, it only detected them at levels well above their optimum usage, making it less efficacious than Intect 7. Clearly, the specific identification of an adulterant is a trade-off for sensitive detection of several adulterants.
机译:评估了三种用于评估用于药物滥用测试的尿液样本完整性的现场掺假检测设备的性能:Intect 7,MASK Ultra Screen和Adultacheck4。Intect7同时检测肌酐,亚硝酸盐,戊二醛, pH值,比重以及漂白剂和吡啶鎓氯铬酸盐(PCC)的存在。 Mask Ultra Screen测试肌酐,亚硝酸盐,pH,比重和氧化剂,而Adultacheck 4测试肌酐,亚硝酸盐,戊二醛和pH。用物质滥用和精神卫生管理局规定的分析物以高于临界浓度50%的浓度制备尿液标本。将隐身,尿液运气,Instant Clean ADD-IT-ive和KLEAR分别添加到添加了药物的尿液样本中,以使其浓度反映出其包装说明书中报告的“最佳”用法,高于或低于该最佳浓度的25%。据报道,隐形是过氧化物酶。人们认为尿毒症是PCC; Instant Clean ADD-it-ive据说含有戊二醛,而Klear是亚硝酸盐。加入以下稀释剂/添加剂的量分别为添加了尿液的尿液的25%,33%和50%:蒸馏水,漂白剂,氨水和醋。在测试的设备中,Intect 7被证明是最敏感的,它正确地表明在所有测试的样品中均存在掺假或稀释剂。为此,必须同时评估所有指示垫。 Adultacheck 4专门评估了尿液完整性的四个特征,因此在检测几种可商购的流行掺假剂的使用方面非常有限。尽管它正确地评估了这四个特征,但它没有检测到使用“隐形”,“尿失禁”或“即时清洁”这一功能。 Mask Ultra Screen可以检测到比Adultacheck 4更广泛的掺假物。但是,实际上,它只能以远高于其最佳用法的水平检测到掺假物,因此其效率不及Intect7。显然,掺假物的具体鉴定是为了敏感地检测几种掺杂物,需要进行权衡。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号