Importance: If not adequately cleaned, rigid nasal endoscopes (RNEs) have the potential to cause iatrogenic cross-contamination. Objective: To test the efficacy of various disinfection methods in reducing bacterial load on RNEs in vitro. Design and Setting: In vitro model. Interventions: Staphylococcus aureus, Streptococcus pneumoniae, Pseudomonas aeruginosa, and Haemophilus influenzae contamination was separately induced on RNEs in vitro. Two experimental sets were completed. The RNEs were disinfected using the following protocols: 30-second scrub with antimicrobial soap (ABS) and water, 30-second scrub with 70% isopropyl alcohol (IA), 30-second scrub with ABS followed by 30-second scrub with IA, 30-second scrub with germicidal cloth, isolated 5-minute soak in an enzymatic soap solution, 5- and 10-minute soaks in ortho-phthalaldehyde, 0.55%, solution (Cidex OPA), and isolated 30-second rinse with tap water, all with 30-second precleaning and postcleaning rinses with tap water. Two sets of experiments (experiment sets A and B) were carried out with a 30-second tap water rinse after inoculation of each RNE. This was followed by immediate cleaning in set A and a 1-hour air-dry delay in set B. Otherwise there were no differences in the disinfection protocols between sets for each method noted. Main Outcomes and Measures: Effectiveness of various disinfection protocols in cleaning rigid nasal endoscopes experimentally inoculated with bacteria commonly found in the upper aerodigestive tract. Positive cultures following disinfection indicated ineffective or incomplete disinfection. Results: Most cleaning methods were effective in eliminating S aureus, S pneumoniae, and H influenzae from the scopes following experimental contamination. Continued growth of P aeruginosa was found after all of the disinfection trials in experiment set A with the exception of a 10-minute immersion in Cidex OPA, and in set B except for the 10-minute Cidex OPA immersion and ABS plus IA trials. Conclusions and Relevance: Most cleaning methods used in our trials appear to properly disinfect RNEs after in vitro inoculation with S aureus, S pneumoniae, and H influenzae. However, it appears that disinfectants may be less effective in cleaning rigid scopes experimentally inoculated with P aeruginosa. There is a paucity of published data regarding cross-contamination during rigid nasal endoscopy, and these results should guide future studies and to some extent practice to avoid iatrogenic spread of contamination.
机译:重要性:如果不充分清洁,刚性鼻内窥镜(RNE)可能会引起医源性交叉污染。目的:测试各种消毒方法在体外减少RNE细菌载量的功效。设计与设置:体外模型。干预措施:体外分别在RNE上诱导了金黄色葡萄球菌,肺炎链球菌,铜绿假单胞菌和流感嗜血杆菌的污染。完成了两个实验组。使用以下协议对RNE进行消毒:使用抗菌肥皂(ABS)和水进行30秒擦洗,使用70%异丙醇(IA)进行30秒擦洗,使用ABS进行30秒擦洗,然后使用IA进行30秒擦洗,用杀菌布擦洗30秒,在酶肥皂溶液中浸泡5分钟,在邻苯二甲醛,0.55%溶液(Cidex OPA)中浸泡5分钟和10分钟,并用自来水冲洗30秒,所有这些均需用自来水进行30秒钟的预清洗和后清洗漂洗。接种每个RNE后,用30秒的自来水冲洗,进行了两组实验(实验组A和B)。随后立即对A组进行清洁,并在B组中进行1小时的风干延迟。否则,对于每种记录的方法,各组之间的消毒方案都没有差异。主要结果和措施:各种消毒方案在清洁刚性鼻内窥镜实验中的有效性,该鼻内窥镜经实验接种了常见于上消化道的细菌。消毒后的阳性培养物表明消毒无效或不完全。结果:大多数清洁方法可有效去除实验污染后范围内的金黄色葡萄球菌,肺炎链球菌和流感嗜血杆菌。除在Cidex OPA中浸泡10分钟和在B组中进行10分钟的Cidex OPA浸泡和ABS加IA试验外,在所有的消毒实验中,实验A中的铜绿假单胞菌均持续生长。结论和相关性:我们的试验中使用的大多数清洁方法似乎在对金黄色葡萄球菌,肺炎链球菌和流感嗜血杆菌进行体外接种后能对RNE进行适当的消毒。但是,看来消毒剂在清洁用铜绿假单胞菌实验接种的刚性内窥镜方面可能不太有效。刚性鼻内窥镜检查期间有关交叉污染的公开数据很少,这些结果应指导将来的研究并在一定程度上避免感染的医源性传播。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号