首页> 外文期刊>Diabetes research and clinical practice >Noninvasive in vitro measurement of pig-blood d-glucose by using a microwave cavity sensor
【24h】

Noninvasive in vitro measurement of pig-blood d-glucose by using a microwave cavity sensor

机译:微波腔传感器无创体外测量猪血d-葡萄糖

获取原文
获取原文并翻译 | 示例

摘要

We have developed an electromagnetic microwave cavity sensor based on the resonant frequency shift for real time measurement of the glycemia in pig blood. We could determine the concentration of d-glucose in pig blood in the range of 150-550mg/dl at the resonance frequency near 4.75GHz with a bandwidth of 300MHz. The change in the d-glucose concentration in blood brings microwave reflection coefficient S 11 changes of about 6.26dB and resonance frequency shifts of about 11.25MHz due to the electromagnetic interaction between the cavity resonator and the blood filled plastic tube inserted into the cavity. This proposed system provides a unique approach for real time noninvasive and contactless glucose monitoring.
机译:我们已经开发了基于共振频率偏移的电磁微波腔传感器,用于实时测量猪血中的血糖。在共振频率为4.75GHz且带宽为300MHz的情况下,我们可以确定猪血中d-葡萄糖的浓度在150-550mg / dl范围内。血液中d-葡萄糖浓度的变化由于空腔谐振器和插入空腔的充血塑料管之间的电磁相互作用,使微波反射系数S 11变化约6.26dB,谐振频率偏移约11.25MHz。该提议的系统为实时无创和非接触式葡萄糖监测提供了独特的方法。