Measurement of the leakage radiation from linear accelerators in the backward direction for 4, 6, 10, 15, and 18 MV x-ray energies.

机译:在4、6、10、15和18 MV X射线能量的反向方向上测量线性加速器的泄漏辐射。

获取原文
获取原文并翻译 | 示例
           

摘要

The x-ray leakage from the housing of a therapy x-ray source is regulated to be <0.1% of the useful beam exposure at 1 m from the source. It is to be expected that the machine leakage in the backward direction would be less because the gantry and stand contain significant amounts of additional metal to attenuate the x rays. A reduction in head leakage in this direction will have a direct effect on the thickness of the shielding wall behind the linear accelerator. However, no reports have been published to date on measurements in this area. The x-ray leakage in the backward direction has been measured from linacs having energies of 4, 6, 10, 15, and 18 MV using a 100 cm ionization chamber and Al2O3 dosimeters. The leakage was measured at nine different positions over the rear wall using a 3 x 3 matrix with a 1-m separation between adjacent horizontal and vertical points with either the leftmost or rightmost column aligned with the target and isocenter. In general, the leakage is less than the canonical value, but the exact value depends on energy, gantry angle, and measurement position. There is significantly greater attenuation directly behind the gantry stand for all energies. Leakage at 10 MV for some positions exceeded 0.1%. Additionally, neutron leakage measurements were made for 10, 15, and 18 MV x-ray beams using track-etch detectors. The average neutron leakage was less than 0.1% except for 18 MV, where neutron leakage was more than 0.1% of the useful beam at some positions.
机译:从治疗X射线源的外壳泄漏的X射线被调节为小于距X射线源1 m处的可用射线曝光量的0.1%。可以预期,由于机架和机架包含大量的额外金属来衰减X射线,因此向后方向的机器泄漏将更少。沿该方向的磁头泄漏减少将对线性加速器后面的屏蔽壁厚度产生直接影响。但是,迄今为止,尚未发布有关该区域测量结果的报告。已使用100 cm电离室和Al2O3剂量计从能量为4、6、10、15和18 MV的直线加速器测量了向后方向的X射线泄漏。使用3 x 3矩阵在后壁的九个不同位置处测量泄漏,相邻水平点和垂直点之间的距离为1-m,最左边或最右边的列与目标和等角点对齐。通常,泄漏量小于规范值,但确切值取决于能量,机架角度和测量位置。所有能量在龙门架后方的衰减明显更大。某些位置的10 MV泄漏超过0.1%。另外,使用轨迹蚀刻检测器对10、15和18 MV X射线束进行了中子泄漏测量。除18 MV外,平均中子泄漏量小于0.1%,其中18 MV在某些位置中子泄漏量大于有用射束的0.1%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号