首页> 外文期刊>Health Physics: Official Journal of the Health Physics Society >Electromagnetic malfunction of semiconductor-type electronic personal dosimeters caused by access control systems for radiation facilities.
【24h】

Electromagnetic malfunction of semiconductor-type electronic personal dosimeters caused by access control systems for radiation facilities.

机译:由辐射设施的访问控制系统引起的半导体型电子个人剂量计的电磁故障。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

High frequency electromagnetic fields in the 120 kHz band emitted from card readers for access control systems in radiation control areas cause abnormally high and erroneous indicated dose readings on semiconductor-type electronic personal dosimeters (SEPDs). All SEPDs malfunctioned but recovered their normal performance by resetting after the exposure ceased. The minimum distances required to prevent electromagnetic interference varied from 5.0 to 38.0 cm. The electric and magnetic immunity levels ranged from 35.1 to 267.6 V m(-1) and from 1.0 to 16.6 A m(-1), respectively. Electromagnetic immunity levels of SEPDs should be strengthened from the standpoint of radiation protection.
机译:从辐射控制区域中的门禁系统的读卡器发出的120 kHz频带中的高频电磁场会在半导体型电子个人剂量计(SEPD)上引起异常高的指示剂量读数。在暴露停止后,所有SEPD均发生故障,但通过重置恢复了正常性能。防止电磁干扰所需的最小距离从5.0到38.0 cm不等。电气和磁抗扰度范围分别为35.1至267.6 V m(-1)和1.0至16.6 A m(-1)。从辐射防护的角度出发,应增强SEPD的电磁抗扰度。