Impact of the Advisa MRI pacing system on the diagnostic quality of cardiac MR images and contraction patterns of cardiac muscle during scans: Advisa MRI randomized clinical multicenter study results.

机译:Advisa MRI起搏系统对扫描过程中心脏MR图像的诊断质量和心肌收缩模式的影响:Advisa MRI随机临床多中心研究结果。

获取原文
获取原文并翻译 | 示例
           

摘要

The Advisa MRI system is designed to safely undergo magnetic resonance imaging (MRI). Its influence on image quality is not well known.To evaluate cardiac magnetic resonance (CMR) image quality and to characterize myocardial contraction patterns by using the Advisa MRI system.In this international trial with 35 participating centers, an Advisa MRI system was implanted in 263 patients. Of those, 177 were randomized to the MRI group and 150 underwent MRI scans at the 9-12-week visit. Left ventricular (LV) and right ventricular (RV) cine long-axis steady-state free precession MR images were graded for quality. Signal loss along the implantable pulse generator and leads was measured. The tagging CMR data quality was assessed as the percentage of trackable tagging points on complementary spatial modulation of magnetization acquisitions (n=16) and segmental circumferential fiber shortening was quantified.Of all cine long-axis steady-state free precession acquisitions, 95% of LV and 98% of RV acquisitions were of diagnostic quality, with 84% and 93%, respectively, being of good or excellent quality. Tagging points were trackable from systole into early diastole (360-648 ms after the R-wave) in all segments. During RV pacing, tagging demonstrated a dyssynchronous contraction pattern, which was not observed in nonpaced (n = 4) and right atrial-paced (n = 8) patients.In the Advisa MRI study, high-quality CMR images for the assessment of cardiac anatomy and function were obtained in most patients with an implantable pacing system. In addition, this study demonstrated the feasibility of acquiring tagging data to study the LV function during pacing.
机译:Advisa MRI系统设计用于安全地进行磁共振成像(MRI)。它对图像质量的影响尚不清楚。使用Advisa MRI系统评估心脏磁共振(CMR)图像质量并表征心肌收缩模式。在这个有35个参与中心的国际试验中,263植入了Advisa MRI系统耐心。在这些患者中,有177名被随机分配到MRI组,并在9-12周就诊时对150名患者进行了MRI扫描。对左心室(LV)和右心室(RV)电影长轴稳态自由进动MR图像进行质量分级。测量沿植入式脉冲发生器和导线的信号损失。标记CMR数据质量的评估方法是:在磁化采集的互补空间调制(n = 16)和分段圆周纤维缩短上可追踪的标记点所占的百分比。在所有电影长轴稳态自由进动采集中,有95% LV和98%的RV采集具有诊断质量,分别为84%和93%的质量为优良。在所有节段中,从收缩期到舒张早期(R波后360-648 ms)均可追踪标记点。在RV起搏过程中,标签显示出不同步的收缩模式,在未起搏(n = 4)和右心房起搏(n = 8)的患者中未观察到。Advisa MRI研究中,用于评估心脏的高质量CMR图像大多数具有植入式起搏系统的患者均获得了解剖学和功能。此外,这项研究证明了获取标记数据以研究起搏期间左室功能的可行性。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号