首页> 外文期刊>Head and neck: Journal for the sciences and specialities of the head and neck >A study of the effect of a resistive heat moisture exchanger (trachinaze) on pulmonary function and blood gas tensions in patients who have undergone a laryngectomy: A randomized control trial of 50 patients studied over a 6-month period.
【24h】

A study of the effect of a resistive heat moisture exchanger (trachinaze) on pulmonary function and blood gas tensions in patients who have undergone a laryngectomy: A randomized control trial of 50 patients studied over a 6-month period.

机译:电阻加热湿气交换剂(trachinaze)对喉切除术后患者肺功能和血气张力的影响的研究:一项为期6个月的50例患者的随机对照试验。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND.: Previous work from this department has shown that resistive tracheostomy filters increase the partial pressure of oxygen in capillary blood and also provide a significant amount of heat/moisture exchange. Until now it has not been shown whether there is any long-term beneficial effect and in particular whether raised tissue oxygenation is maintained using a practical filter device. METHODS.: We carried out a 6-month randomized control trial including 50 laryngectomees. Twenty-five patients were treated with the Liverpool Heat Moisture Exchange device incorporating an airway resistor (Trachinaze). Another 25 patients were treated with a placebo device. Relevant subjective and objective data were collected before and at the end of the study. The objective measurements were capillary oxygen tension (which parallels blood arterial tension), carbon dioxide tension, FEV(1), FVC, and PIF. Patients were reviewed at intervals throughout the study. Data were analyzed using the Mann-Whitney U test and the paired t test to test the difference between the active device and placebo at 6 months. RESULTS.: Subjective lower airway parameters, including cough, number of chest infections, mucus production, and shortness of breath at rest, were significantly improved in the active group compared with the placebo group. The objective parameters FEV(1), FVC, and PIF were not significantly different. Capillary oxygen tension, however, was highly significantly raised in the active group at 6 months. CONCLUSIONS.: Trachinaze is highly superior to placebo at improving subjective pulmonary parameters, including shortness of breath. It is also superior in its ability to maintain an increased peripheral tissue oxygen tension over a 6-month period.
机译:背景:该部门的先前工作表明,电阻式气管切开过滤器会增加毛细血管血液中氧气的分压,并提供大量的热量/水分交换。迄今为止,尚未显示出是否存在任何长期的有益效果,特别是使用实用的过滤器装置是否能够维持升高的组织氧合作用。方法:我们进行了为期6个月的随机对照试验,其中包括50个喉癌切除组。 25例患者接受了利物浦热气交换装置的治疗,该装置装有导气管电阻器(Trachinaze)。另有25名患者接受了安慰剂治疗。在研究之前和结束时收集相关的主观和客观数据。客观测量为毛细血管氧张力(与血液动脉张力平行),二氧化碳张力,FEV(1),FVC和PIF。在整个研究过程中定期检查患者。使用Mann-Whitney U检验和配对t检验对数据进行分析,以测试6个月时有源设备和安慰剂之间的差异。结果:与安慰剂组相比,活动组的主观下呼吸道参数(包括咳嗽,胸部感染次数,粘液产生和静息气短)明显改善。客观参数FEV(1),FVC和PIF没有显着差异。然而,活动组在6个月时毛细血管氧张力显着升高。结论:曲沙霉素在改善包括呼吸急促在内的主观肺部参数方面,优于安慰剂。它在六个月内维持增加的外周组织氧张力的能力也很出色。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号