首页> 外文期刊>Vascular medicine >Reproducibility of different methods to measure the endothelial function
【24h】

Reproducibility of different methods to measure the endothelial function

机译:不同方法测量内皮功能的重现性

获取原文
获取原文并翻译 | 示例
           

摘要

This study compares the reliability and reproducibility of flow-mediated dilatation (FMD) and peripheral arterial tonometry (PAT) to assess endothelial function. Eighteen volunteers with coronary heart disease underwent simultaneous testing of FMD by means of brachial artery ultrasound scanning and PAT with the EndoPAT 2000. The subjects underwent both tests twice with an interval of 30 minutes and the same protocol was repeated 2 days later. The within-day variability was lower for the FMD measurements than for the PAT measurements (10% versus 18%; p < 0.05). The between-day variability was similar (11%). A significant correlation was found between PAT and FMD (r = 0.57, p < 0.001). The typical errors of measurements for FMD were lower than those for PAT. In conclusion, in this study, FMD demonstrated the best reliability for the within- and between-day and separate day measurements when compared to PAT.
机译:本研究比较了血流介导的扩张(FMD)和外周动脉眼压(PAT)的可靠性和可重复性,以评估内皮功能。 18名患有冠心病的志愿者通过肱动脉超声扫描和PAT与EndoPAT 2000同时进行了FMD测试。受试者均以30分钟的间隔两次进行了两次测试,并在2天后重复了相同的方案。 FMD测量的日内变异性低于PAT测量的变异(10%比18%; p <0.05)。日间变异性相似(11%)。在PAT和FMD之间发现显着相关性(r = 0.57,p <0.001)。 FMD的典型测量误差低于PAT。总之,在这项研究中,与PAT相比,FMD证明了日内和日间以及日间测量的最佳可靠性。