首页> 外文期刊>Transfusion: The Journal of the American Association of Blood Banks >Effect on the quality of blood components after simulated blood transfusions using volumetric infusion pumps.
【24h】

Effect on the quality of blood components after simulated blood transfusions using volumetric infusion pumps.

机译:使用容积式输液泵模拟输血后对血液成分质量的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: It is unknown whether the use of volumetric infusion pumps for the transfusion of red blood cells (RBCs) or platelet (PLT) concentrates (PCs) affects the quality of the blood components. We therefore investigated the in vitro quality of these components after use of infusion pumps. STUDY DESIGN AND METHODS: Ten different volumetric infusion pumps were used to simulate transfusion with RBCs and PCs. To prevent donor-dependent differences multiple units were pooled and divided into equal portions. The storage time of RBCs was 30 to 35 days (n=10 experiments), and for PCs, either 2 (n=5) or 7 days (n=5). For RBCs an infusion rate of 100 or 300mL/hr was used, and for PCs, 600mL/hr. Transfusions without an infusion pump served as a reference. RESULTS: None of the infusion pumps induced an increase of free hemoglobin, annexin A5 binding, or formation of echinocytes in RBCs compared to reference units. In 2- and 7-day-old PCs no effect was shown on PLT concentration, annexin A5 binding, mean PLT volume, and morphology score compared to the reference. The CD62P expression of 2-day-old PCs was significantly lower after transfusion compared to the reference, that is, 11.7+/-2.1% versus 8.1+/-1.3% (p<0.01). CONCLUSION: There was no adverse effect on the in vitro quality of RBCs or PCs after simulated transfusion using volumetric infusion pumps. A decrease in PLT activation was observed, which can probably be explained by capturing of activated or damaged PLTs in the 200-microm filter present in the infusion system.
机译:背景:尚不清楚使用容积式输液泵输血红细胞(RBC)或血小板(PLT)浓缩液(PC)是否会影响血液成分的质量。因此,我们在使用输液泵后调查了这些组件的体外质量。研究设计和方法:十个不同的容量输液泵用于模拟RBC和PC的输血。为防止供体依赖性差异,将多个单元合并,并分成相等的部分。 RBC的存储时间为30到35天(n = 10个实验),而PC的存储时间为2天(n = 5)或7天(n = 5)。对于RBC,使用的输液速度为100或300mL / hr,对于PC,使用的输液速度为600mL / hr。没有输液泵的输血作为参考。结果:与参考单位相比,所有输液泵均未引起红细胞中游离血红蛋白,膜联蛋白A5结合或棘突细胞形成的增加。与参考相比,在2天和7天龄的PC中,对PLT浓度,膜联蛋白A5结合,平均PLT体积和形态评分均无影响。与参考相比,输血后2天龄PCs的CD62P表达明显降低,即分别为11.7 +/- 2.1%和8.1 +/- 1.3%(p <0.01)。结论:使用体积输液泵模拟输注后,对RBC或PC的体外质量没有不利影响。观察到PLT激活减少,这可能是由于在输液系统中存在的200微米过滤器中捕获了激活或损坏的PLT所致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号