When samples are analysed with elastic recoil detection analysis (ERDA) there are common limitations due to properties of the sample or the experimental set-up. Typical problems originate from the decomposition or the roughness of the sample, from the resolution that can be achieved with the detector system or from the accuracy of the ion stopping power values. Nevertheless, there are different experimental and evaluation techniques that can be applied to avoid such problems or to deal with measurements of difficult samples. The detection system can be improved by dedicated evaluation procedures. Different strategies are currently used to improve the examination of ERDA spectra and the validity of the results obtained. Copyright (C) 2003 John Wiley Sons, Ltd. [References: 8]
机译:当使用弹性反冲检测分析(ERDA)分析样品时,由于样品的性质或实验装置的原因,存在共同的局限性。典型问题源自样品的分解或粗糙度,检测器系统可实现的分辨率或离子终止功率值的准确性。但是,可以应用各种不同的实验和评估技术来避免此类问题或处理困难样品的测量。可以通过专用评估程序来改进检测系统。当前使用不同的策略来改善ERDA光谱的检查以及所获得结果的有效性。版权所有(C)2003 John Wiley Sons,Ltd. [参考:8]