Evaluation of a new set of automated chemiluminescense assays for anticardiolipin and anti-beta2-glycoprotein i antibodies in the laboratory diagnosis of the antiphospholipid syndrome

机译:评价抗心磷脂和抗β2-糖蛋白i抗体的一组新的自动化化学发光测定在抗磷脂综合征实验室诊断中的价值

获取原文
获取原文并翻译 | 示例
           

摘要

Introduction: The laboratory diagnosis of antiphospholipid syndrome (APS) requires the demonstration of antiphospholipid antibodies (aPL): lupus anticoagulant (LAC) measured through coagulation assays, anticardiolipin IgG or IgM antibodies (aCL) and/or anti-β2glycoprotein I IgG or IgM antibodies (aβ2GPI), usually detected by ELISA Materials and methods: We evaluated the diagnostic value of aCL and aβ2GPI measured by a new automated system using the chemiluminescence principle, the immunoanalyzer Zenit RA (Menarini). Results: Results of aCL and aβ2GPI were correlated with the clinical background of the patients and with results of ELISA (n = 314). Correlated to the clinical background sensitivity/specificity ranged for aCL IgG between 7.5-45.2% / 54.2-98.8%, for aCL IgM 3.4-5.5% / 89.9-94%, for aβ2GPI IgG 5.5-25.3% / 75.6-100% and aβ2GPI IgM 3.4-4.8% / 89.9-92.3%, depending on the cut-off used. Sensitivity with manufacturer's cut-offs was comparable to ELISA, except for aβ2GPI IgG with a significantly lower sensitivity compared to ELISA (5.5% vs 11.6%). In the APS patient population (n = 30) sensitivity of aCL IgG and aβ2GPI IgG was higher measured by ELISA compared to Zenit RA (46.7% vs 30.0%, and 46.7% vs 26.7%, respectively). Agreement between Zenit RA results and ELISA results for the four parameters was moderate (Kappa-values ranging 0.509-0.565). Sensitivity was 38.5%, 53.3%, 40% and 69.2% for aCL IgG, aCL IgM, aβ2GPI IgG and aβ2GPI IgM, respectively, applying the highest cut-off value for Zenit RA, raising towards 64.3%, 100%, 57.1%, for aCL IgG, aCL IgM, aβ2GPI IgG, respectively, in a APS patient population. Conclusions: The new technology of chemiluminescense for measuring aPL showed good performance characteristics. Interpretation of results with a cut-off value associated with a good discrimination for disease, resulted in a lower sensitivity for the diagnosis of APS for aβ2GPI IgG measured by Zenit RA assays compared to ELISA; sensitivity for aCL IgG was comparable to ELISA. Specificity for all parameters was high and comparable for aCL and aβ2GPI.
机译:简介:抗磷脂综合征(APS)的实验室诊断需要证明抗磷脂抗体(aPL):通过凝血测定,抗心磷脂IgG或IgM抗体(aCL)和/或抗β2糖蛋白I IgG或IgM抗体测定的狼疮抗凝剂(LAC) (aβ2GPI),通常通过ELISA进行检测。材料和方法:我们使用化学发光原理,免疫分析仪Zenit RA(Menarini),通过新型自动化系统评估了aCL和aβ2GPI的诊断价值。结果:aCL和aβ2GPI的结果与患者的临床背景和ELISA的结果相关(n = 314)。与临床背景敏感性/特异性相关的aCL IgG介于7.5-45.2%/ 54.2-98.8%,aCL IgM 3.4-5.5%/ 89.9-94%,aβ2GPIIgG 5.5-25.3%/ 75.6-100%和aβ2GPI IgM 3.4-4.8%/ 89.9-92.3%,取决于使用的临界值。制造商临界值的灵敏度与ELISA相当,只是aβ2GPIIgG的灵敏度比ELISA低得多(5.5%比11.6%)。在APS患者群体(n = 30)中,通过ELISA测定的aCL IgG和aβ2GPIIgG的敏感性高于Zenit RA(分别为46.7%和30.0%,和46.7%和26.7%)。四个参数的Zenit RA结果与ELISA结果之间的一致性中等(Kappa值在0.509-0.565之间)。对aCL IgG,aCL IgM,aβ2GPIIgG和aβ2GPIIgM的敏感性分别为38.5%,53.3%,40%和69.2%,应用Zenit RA的最高临界值,分别提高到64.3%,100%,57.1%,在APS患者群体中分别检测aCL IgG,aCL IgM和aβ2GPIIgG。结论:化学发光法测定aPL的新技术具有良好的性能特点。与酶联免疫吸附测定相比,对结果的临界值的解释与对疾病的良好区分相关,因此对APS诊断aβ2GPIIgG的敏感性较低。对aCL IgG的敏感性与ELISA相当。所有参数的特异性都很高,与aCL和aβ2GPI相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号