Evaluation of four portable blood glucose meters in diabetic and non-diabetic dogs and cats

机译:糖尿病和非糖尿病犬猫中四种便携式血糖仪的评估

获取原文
获取原文并翻译 | 示例
       

摘要

Background: Monitoring of an animal's blood glucose concentration is critical for diagnostic and therapeutic decisions. Over the past few decades, portable blood glucose meters (PBGMs) have been used to monitor blood glucose concentrations in animals. Recently, new and improved PBGMs have been made available on the market.Objective: The purpose of this study was to evaluate four PBGMs for use in dogs and cats.Animals and methods: A total of 155 venous blood samples of dogs and 85 venous blood samples of cats were tested using four PBGMs. Control solutions from manufacturers were used to determine the precision of each meter. The coefficient of variation was calculated to determine precision during a set of replicates. Pearson's correlation analysis, Passing-Bablok regression, and Bland-Altman analysis were used to determine the accuracy of four PBGMs against the hexokinase reference method. Error grid analysis was used to evaluate clinical relevance.Results: All PBGMs, except CERA-PET (R), were clinically acceptable for monitoring blood glucose concentrations; AlphaTrak (R) and VetMate (R) appeared to be the most accurate ones, demonstrating that to use PBGMs for glucose monitoring, it is important to understand the strengths or limitations of each meter. The difference in results between the PBGMs and the reference method increased at high glucose concentration ranges, which were also affected by the hematocrit.Conclusions: Although readings of the PBGMs and the reference method varied across glycemic ranges (low, normal, and high glucose concentrations), most PBGMs were clinically acceptable for monitoring blood glucose concentrations in dogs and cats.
机译:背景:监测动物的血糖浓度对于诊断和治疗决策至关重要。在过去的几十年中,便携式血糖仪(PBGM)已用于监测动物中的血糖浓度。最近,市场上出现了新的和改良的PBGMs。目的:本研究的目的是评估用于狗和猫的四种PBGMs。动物和方法:总共155种狗静脉血和85种静脉血使用四个PBGM测试了猫的样本。使用制造商的控制解决方案来确定每个仪表的精度。计算变异系数以确定一组重复过程中的精度。使用Pearson的相关分析,Passing-Bablok回归和Bland-Altman分析来确定四种PBGM相对于己糖激酶参考方法的准确性。结果:所有除CERA-PET(R)以外的PBGM在临床上均可接受用于监测血糖浓度。 AlphaTrak(R)和VetMate(R)似乎是最准确的,这表明要使用PBGM进行葡萄糖监测,了解每个仪表的优势或局限性很重要。 PBGMs和参考方法之间的结果差异在高葡萄糖浓度范围内有所增加,这也受到血细胞比容的影响。结论:尽管PBGMs和参考方法的读数在血糖范围内(低,正常和高葡萄糖浓度)有所不同),大多数PBGM在临床上都可以用于监测狗和猫的血糖浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号