OBJECTIVES/HYPOTHESIS: To compare the hearing outcomes and complications observed using either titanium or nontitanium prostheses in a 7-year consecutive series of ossiculoplasties performed by a single surgeon. STUDY DESIGN: Retrospective. METHODS:: A database of ossicular reconstruction surgeries was reviewed for preoperative and postoperative audiometric data including air and bone conduction thresholds at four frequencies and speech reception thresholds. Outcomes were evaluated at time points less than and greater than 6 months postoperatively. Baseline demographic and surgical characteristics and postoperative complications were also noted. RESULTS: A total of 105 cases had sufficient audiometric data available for analysis, including 80 performed with titanium and 25 with nontitanium implants. Follow-up ranged from 1.2 to 74.2 months, with a mean of 14.9 months. Mean air-bone gap at initial follow-up was 21.7 dB in the nontitanium group and 15.4 dB in the titanium group; this difference was significant (P = .01). Postoperative air-bone gap of less than 20 dB at initial follow-up was achieved in 50.0% of nontitanium cases and 77.1% of titanium cases (P = .012). This difference in "success" rates persisted at longer follow-up but did not achieve statistical significance. Mean speech reception thresholds at <6 months was 29.7 dB in the nontitanium group and 22.6 dB in the titanium group (P = .049). Extrusion was observed with two nontitanium prostheses (8.0%) and three titanium prostheses (3.8%) (P > .05). CONCLUSIONS: Titanium ossicular prostheses provide hearing outcomes superior to those of nontitanium prostheses when evaluated within 6 months after ossiculoplasty.
机译:目的/假设:为了比较在由一名外科医生进行的连续7年的一系列眼球成形术中,使用钛或非钛假体观察到的听力结果和并发症。研究设计:回顾性研究。方法:审查了听骨重建手术的数据库,以获取术前和术后听觉测验数据,包括四个频率的空气和骨传导阈值以及语音接收阈值。在术后少于和大于6个月的时间点评估结果。还注意到基线的人口统计学和手术特征以及术后并发症。结果:总共105例具有足够的听力数据可用于分析,其中80例使用钛,而25例使用非钛植入物。随访时间为1.2至74.2个月,平均14.9个月。初次随访时,非钛组的平均气隙为21.7 dB,钛组的平均气隙为15.4 dB。这种差异是显着的(P = 0.01)。 50.0%的非钛病例和77.1%的钛病例在初次随访时的术后气隙小于20 dB(P = 0.012)。随访时间较长时,“成功”率的这种差异持续存在,但未达到统计学意义。 <6个月时,非钛组的平均语音接收阈值为29.7 dB,钛组为22.6 dB(P = .049)。观察到有两个非钛假体(8.0%)和三个钛假体(3.8%)的挤压(P> .05)。结论:在听骨成形术后6个月内进行评估时,钛听骨假体的听觉效果优于非钛听骨假体。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号