Reliability and sensitivity of joint space measurements in hand radiographs using computerized image analysis.

机译:使用计算机图像分析的手部X光片中关节间隙测量的可靠性和敏感性。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To establish the sensitivity and reliability of proximal interphalangeal (PIP) and metacar pophalangeal (MCP) mean joint space measurements using standard clinical radiographs of healthy subjects, in order to determine the limits at which a change in radiographic joint space could indicate a change in actual joint size. METHODS: Repeat hand radiographs of healthy subjects were taken using standard techniques at 3-5 day intervals with the hands flat (5 posteroanterior radiographs in 8 subjects) or in 6 different flexed positions on a single occasion (8 subjects). The mean joint space was determined using custom soft ware and was validated manually. Measurement reproducibility within subjects, within films, and between hand positions was assessed by analysis of variance. RESULTS: In repeat radiographs taken in the standard clinical position, the precision of individual join space measurements indicates that changes > 0.11 mm (approximately 7%) would represent an actual physica change in joint space width (with 95% probability). Averaging measurements across fingers for a single subject decreases the detectable change to 0.05 mm (approximately 3%). With increasing flexure, radiographic joint space tended to increase in MCP and decrease in PIP. CONCLUSION: Mean finger joint space measured from standard clinical radiographs is a reliable and sensitive measurement in healthy subjects even with some change in hand position. Work is required to establish whether the joint space change measured from serial radiographs of patients with arthritis over a period of 6-12 mo exceeds the detectable limits of change derived in this study.
机译:目的:利用健康受试者的标准临床X线照片确定近端指间(PIP)和meta骨后hal(MCP)平均关节间隙的敏感性和可靠性,以确定放射线关节间隙改变可指示改变的极限实际的关节尺寸。方法:使用标准技术以3-5天的间隔重复拍摄健康受试者的手部X线照片,双手放平(8位受试者5次后前位X线照片),或单次在6个不同屈曲位置(8位受试者)进行摄影。平均关节间隙使用定制软件确定并手动验证。通过方差分析评估对象内部,胶片内部以及手部位置之间的测量重现性。结果:在标准临床位置拍摄的重复X射线照片中,单个关节间隙测量的精度表明,变化> 0.11毫米(约7%)将代表关节间隙宽度的实际生理变化(可能性为95%)。单个对象的手指平均测量将可检测到的变化减少到0.05毫米(约3%)。随着挠度的增加,MCP的放射线关节间隙趋于增大,而PIP的放射线关节间隙趋于减小。结论:根据标准的临床X光片测量的平均手指关节间隙是健康受试者的可靠且敏感的测量结果,即使手部位置有所变化。需要开展工作来确定在6至12个月的时间内从关节炎患者的X线照片测量的关节间隙变化是否超出了本研究得出的可检测到的变化极限。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号