首页> 外文期刊>The Journal of the American Dental Association >Assessing the irradiance delivered from light-curing units in private dental offices in Jordan
【24h】

Assessing the irradiance delivered from light-curing units in private dental offices in Jordan

机译:评估约旦私人牙科诊所的光固化机发出的辐照度

获取原文
获取原文并翻译 | 示例
           

摘要

Background. The authors conducted a study to examine the irradiance from light-curing units (LCUs) used in dental offices in Jordan. Methods. Two of the authors visited 295 private dental offices (15 percent) in Jordan and collected the following information about the LCUs: age, type (quartztungsten- halogen or light-emitting diode), date of last maintenance, type of maintenance, last date of use, number of times used during the day, availability of a radiometer, exposure time for each resin-based composite increment, size of light-curing tips and presence of resin-based composite on the tips. The authors used a radiometer to measure the irradiance from the LCUs. They used linear regression with stepwise correlation for the statistical analysis. The authors set the minimum acceptable irradiance at 300 milliwatts/square centimeter. Results. The mean irradiance of the 295 LCUs examined was 361 mW/cm2, and 136 LCUs (46.1 percent) delivered an irradiance of less than 300 mW/cm2. The unit's age, type and presence of resin-based composite on the light-curing tips had a significant effect on the irradiance (P ≤ .001). Conclusions. Only 37 of the 141 quartz-tungsten-halogen units (26.2 percent) and 122 of the 154 light-emitting diode units (79.2 percent) delivered at least 300 mW/cm 2. Resin contamination on the light-curing tips had a significant effect on the irradiance delivered. The irradiance from the LCUs decreased with use. Practical Implications. The irradiance from many of the units in this study was less than 300 mW/cm2, which may affect the quality of resin-based composite restorations. Dentists should monitor the performance of the LCUs in their offices weekly.
机译:背景。作者进行了一项研究,以检查约旦牙科诊所使用的光固化单元(LCU)的辐射。方法。其中两位作者访问了约旦的295家私人牙科诊所(占15%),并收集了以下有关LCU的信息:年龄,类型(石英钨-卤素灯或发光二极管),最后一次维护的日期,维护的类型,最后一次维护的日期。使用,一天中使用的次数,辐射计的可用性,每种树脂基复合材料增量的暴露时间,光固化笔尖的尺寸以及在笔尖上存在树脂基复合材料。作者使用辐射计测量来自LCU的辐照度。他们使用具有逐步相关性的线性回归进行统计分析。作者将最小可接受辐照度设置为300毫瓦/平方厘米。结果。被检查的295个LCU的平均辐照度为361 mW / cm2,而136个LCU(46.1%)的辐照度小于300 mW / cm2。单位的年龄,类型和光固化尖端上树脂基复合材料的存在对辐照度有显着影响(P≤.001)。结论。 141个石英-钨-卤素单元中只有37个(26.2%)和154个发光二极管单元中的122个(79.2%)的输出功率至少为300 mW / cm2。光固化头上的树脂污染产生了显着影响在提供的辐照度上。 LCU的辐照度随着使用而降低。实际影响。本研究中许多单元的辐照度小于300 mW / cm2,这可能会影响树脂基复合修复体的质量。牙医应每周监控办公室中LCU的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号