Assessment of cold welding properties of the internal conical interface of two commercially available implant systems.

机译:评估两个商用植入系统的内部圆锥形界面的冷焊性能。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

STATEMENT OF PROBLEM: The cone-screw abutment has been shown to diminish micromovement by reducing the burden of component loosening and fracture. However, anecdotal concern for cold welding of cone-screw joints in implant design has been identified as a potential source for lack of retrievability. PURPOSE: This comparative study evaluated the loosening torque, as a percentage of tightening torque, for the ITI Straumann and Astra Tech (3.5 and 4.0 mm diameters) implant systems, which use an 8-degree and 11-degree internal cone, respectively. MATERIAL AND METHODS: Implants and abutments from each system were mounted in a torque device, and a range of tightening torques was applied. Loosening torques were then measured, and the influence of conus angle, interfacial surface area, saliva contamination, and time delay to loosening were all assessed. RESULTS: The loosening torque only exceeded tightening torque at the highest levels, just before component failure, when plastic deformation was expected. For all clinically relevant levels of torque, both in a dry environment and with components bathed in artificial saliva at 37 degrees C, loosening torque was always seen to be 80% to 90% of tightening torque, demonstrating that cold welding does not occur. There was a high correlation between loosening and tightening torque for all systems tested, but no statistical difference when comparing wet versus dry or comparing individual data for each system. CONCLUSIONS: It can be concluded that for clinically relevant levels of tightening torque, no problems are anticipated with respect to retrievability.
机译:问题陈述:通过减少零件松动和断裂的负担,锥形螺丝基台已被证明可减少微动。然而,在种植体设计中,锥螺纹接头冷焊的传闻已被认为是缺乏可回收性的潜在原因。目的:该比较研究评估了分别使用8度和11度内锥的ITI Straumann和Astra Tech(直径为3.5毫米和4.0毫米)植入系统的松动扭矩,以拧紧扭矩的百分比表示。材料和方法:将每个系统的植入物和基台安装在扭矩装置中,并施加一定范围的拧紧扭矩。然后测量松动扭矩,并评估锥角,界面表面积,唾液污染和松开时间延迟的影响。结果:在预期发生塑性变形时,在组件失效之前,松动扭矩仅在最高水平超过了拧紧扭矩。对于所有在临床上相关的扭矩水平,无论是在干燥环境中还是将组件浸入37摄氏度的人工唾液中,松动扭矩始终被视为紧固扭矩的80%至90%,这表明未发生冷焊。在所有测试的系统中,松动扭矩和拧紧扭矩之间具有高度相关性,但在比较干湿两用或比较每个系统的单个数据时,没有统计学差异。结论:可以得出结论,对于临床上相关的拧紧扭矩水平,可恢复性没有问题。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号