To improve the detection sensitivity of metal ions in capillary electrophoresis (CE), combination of on-line complexation and sweeping was explored using cyclokexane-1, 2-diaminetetraacetic acid (CDTA). CDTA has two roles, as a dynamic complexation agent to form a UV-absorbing complex and a carrier for sweeping. Six metal ions were used as test analytes. The effect of pH, CDTA concentration, composition of BGS and sample matrix, injection length of sample solution were examined and optimized to obtain high performance of this method. Under optimized conditions, the technique was validated in terms of the limit of detection, repeatability, and sensitivity enhancement. improved detection responses were 60- to 160-Fold in terms of peak heights for metal ions in comparison with conventional CZE of precapillary complexed metal ions. The limits of detection were in the range of 0.4 - 4.2 x 10(-7) M.
机译:为了提高毛细管电泳(CE)中金属离子的检测灵敏度,使用环酮1-1、2-二胺四乙酸(CDTA)探索了在线络合和扫描的结合。 CDTA具有两个作用,作为形成紫外线吸收复合物的动态络合剂和清扫载体。六种金属离子用作测试分析物。检测并优化了pH,CDTA浓度,BGS的组成和样品基质的影响,样品溶液的进样长度,从而获得了该方法的高性能。在优化条件下,该技术在检测限,可重复性和灵敏度增强方面得到了验证。与常规CZE的毛细管前复合金属离子相比,金属离子的峰高改善了检测响应,提高了60到160倍。检测限在0.4-4.2 x 10(-7)M的范围内。