Comparison of ACL fixation devices using cadaveric grafts.

机译:使用尸体移植的ACL固定装置的比较。

获取原文
获取原文并翻译 | 示例
       

摘要

We evaluated two newer forms of femoral fixation of hamstring grafts for anterior cruciate ligament reconstruction, the Endobutton direct (Smith and Nephew, Andover, MA) and Femoral intrafix (Depuy Mitek, Raynham, MA), and compare them to devices that have been evaluated in the literature, the AXL Crosspin (Biomet, Warsaw, IN) and Biotransfix II (Arthrex, Naples, FL). Paired hamstring tendon allografts were fixed in the femoral tunnel of 24 cadaveric bovine knees (6 per group) according to each device's specifications. The free ends (tibial sides) were fixed to the materials testing machine via custom-made cryo-clamps. In Phase I, single load to failure and stiffness were evaluated, and in Phase II, peak displacement was evaluated while cyclic loading was performed over 1000 cycles. One-way analyses of variance were performed to test for differences between groups. There were no significant differences in failure load (p = 0.42) or stiffness (p = 0.39) between the fixation devices. There was also no significant difference in peak displacement measured during the cyclic loading phase (p = 0.32). Our findings suggest that the newer generation devices, Endobutton direct and Femoral intrafix, have similar strength in single load to failure and similar peak displacement during cyclic loading as compared with clinically proven Crosspin techniques. These newer devices, which are designed to accommodate for more anatomic femoral tunnel placement, may provide a reasonable alternative without compromising biomechanical properties.
机译:我们评估了用于前交叉韧带重建的两种较新形式的股筋固定术,即Endobutton Direct(马萨诸塞州安德沃市的Smith和Nephew)和股骨内固定器(Depuy Mitek,Raynham,马萨诸塞州),并将它们与已评估的装置进行了比较在文献中,AXL Crosspin(Biomet,Warsaw,IN)和Biotransfix II(Arthrex,Naples,FL)。根据每种器械的规格,将成对的绳肌腱同种异体移植物固定在24具尸体牛膝盖的股骨隧道中(每组6个)。自由端(胫骨侧)通过定制的冷冻夹固定在材料测试机上。在阶段I中,评估了单个破坏载荷和刚度,在阶段II中,评估了峰值位移,同时进行了1000次以上的循环载荷。进行单向方差分析以检验组之间的差异。固定装置之间的破坏载荷(p = 0.42)或刚度(p = 0.39)没有显着差异。在循环加载阶段测得的峰位移也没有显着差异(p = 0.32)。我们的发现表明,与临床验证的Crosspin技术相比,Endobutton Direct和股骨内固定装置等新一代装置在单载荷至失效时具有相似的强度,在循环载荷过程中具有相似的峰值位移。这些设计用于适应更多解剖性股骨隧道放置的设备可以在不损害生物力学性能的情况下提供合理的替代方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号