The primary goal of this study was to determine the initial press-fit stability in acetabular components without screw fixation. Mechanical testing was performed with the implantation of press-fit acetabular components in cadaveric specimens. No significant difference was found in load to failure testing between 1 and 2. mm of under-reaming. However, there was significant variability in bending forces required to create 150. μm of micromotion ranging from 49.3. N to 214.4. N. This study shows that cups implanted in a press-fit fashion, which are felt to be clinically stable, have high degrees of variability in resisting load and may be at risk for loosening. There is a need for more objective intra-operative techniques to test cup stability.
机译:这项研究的主要目的是确定没有螺钉固定的髋臼组件的初始压入配合稳定性。在尸体标本中植入压入式髋臼组件进行机械测试。在欠载荷测试中,欠扩孔1到2 mm之间没有发现显着差异。但是,产生150.μm的微动所需的弯曲力存在很大差异,范围为49.3。 N为214.4。 N.这项研究表明,以压入配合方式植入的杯子在临床上被认为是稳定的,在抵抗负荷方面具有高度的可变性,并且可能有松动的风险。需要更客观的术中技术来测试杯的稳定性。