Knee-simulator testing of conventional and cross-linked polyethylene tibial inserts.

机译:传统和交联聚乙烯胫骨插入物的膝部模拟器测试。

获取原文
获取原文并翻译 | 示例
           

摘要

We compared the resistance to delamination and to adhesive/abrasive wear of conventional and highly cross-linked polyethylene tibial inserts of a cruciate-retaining total knee design using a knee simulator. Both groups were tested after aggressive, accelerated aging, and 1 set of conventional inserts was studied without aging. Aging oxidized the conventional, but not the highly cross-linked, inserts. The simulated normal gait testing lasted for 5 and 10 million cycles for the conventional and highly cross-linked inserts, respectively. Aged conventional inserts showed delaminations, whereas none were observed in the unaged conventional and aged cross-linked inserts. Wear rates measured by the gravimetric method were 9 +/- 2 mm(3), 10 +/- 4 mm(3), and 1 +/- 0 mm(3) per million cycles; by the metrologic method, they were 8 +/- 1 mm(3), 9 +/- 2 mm(3), and 3 +/- 0 mm(3) for the unaged conventional, aged conventional, and aged highly crosslinked inserts, respectively. In the test model used, oxidation led todelamination, whereas increased cross-link density resulted in reduced adhesive/abrasive wear of tibial inserts.
机译:我们比较了采用膝盖模拟器的十字形固定全膝设计的传统和高度交联的聚乙烯胫骨胫骨插入物的抗分层性和粘附/磨蚀性。两组均在积极,加速老化后进行了测试,并研究了1组常规刀片未老化的情况。时效会氧化传统的而非高交联度的刀片。对于常规的和高度交联的插入物,模拟的正常步态测试分别进行了5和1000万个循环。老化的常规刀片显示分层,而在未老化的常规和老化的交联刀片中未观察到分层。重量法测量的磨损率为每百万个循环9 +/- 2 mm(3),10 +/- 4 mm(3)和1 +/- 0 mm(3);通过计量方法,对于未老化的常规,老化的常规和老化的高度交联刀片,它们分别为8 +/- 1 mm(3),9 +/- 2 mm(3)和3 +/- 0 mm(3) , 分别。在使用的测试模型中,氧化导致脱层,而增加的交联密度导致胫骨插入物的粘合/磨料磨损减少。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号