We report the accuracy of positioning of the calibration ball in the process of pre-operative templating for total hip arthroplasty (THA). The ball should be placed in the coronal plane of the hip to provide suitable accuracy. We reviewed 112 post-operative THA radiographs where a calibration ball had been placed. We templated the femoral head size of the implant after calibrating the templating system from the ball. The calibrated femoral head diameter was compared to the known prosthetic head size. A percentage error was calculated. Overall, incorrect placement of the calibration ball resulted in a mean percentage error in templating of 6.8% (median 5.7%; range 0-26%). Such error carries implications with the templating process and may result in incorrect component sizes, leg lengths and offset.
机译:我们报告在全髋关节置换术(THA)术前模板制作过程中校准球的定位精度。球应放置在臀部的冠状平面内,以提供适当的准确性。我们回顾了112例放置了校准球的THA术后X射线照片。在从球上校准模板系统后,我们对植入物的股骨头尺寸进行模板化。将校准后的股骨头直径与已知的假体头部大小进行比较。计算百分比误差。总体而言,错误放置校准球会导致模板中的平均百分比误差为6.8%(中位数为5.7%;范围为0-26%)。这种错误会牵涉到模板过程,并可能导致组件尺寸,支脚长度和偏移量不正确。