首页> 外文期刊>The Journal of arthroplasty >Component Size Mismatch of Metal on Metal Hip Arthroplasty: An Avoidable Never Event
【24h】

Component Size Mismatch of Metal on Metal Hip Arthroplasty: An Avoidable Never Event

机译:金属在髋关节置换术上的部件尺寸不匹配:避免可避免的事件

获取原文
获取原文并翻译 | 示例
           

摘要

Size mismatch of components used in total hip arthroplasty is a serious, preventable patient safety incident of unknown prevalence as many cases are not detected. Component size mismatch was found in 11 cases (0.9%) at our retrieval centre. All cases of mismatch were not detected on plain radiograph during routine clinical follow up and blood metal ion levels were elevated above the MHRA action level of 7. ppb. Root cause analysis identified manufacturer, hospital and surgeon factors that need to be addressed to reduce the incidence of this avoidable clinical problem. Retrieval analysis is the only method of confirming size mismatch and is likely to be under-represented in National Joint Registries that record the indication for revision at the time of revision.
机译:全髋关节置换术中使用的组件尺寸不匹配是一种严重的,可预防的,未知患病率的患者安全事件,因为未检测到许多病例。在我们的检索中心发现11例(0.9%)组件大小不匹配。在常规临床随访期间,未在平片上检测到所有错配病例,血液金属离子水平升高至MHRA作用水平7 ppb以上。根本原因分析确定了制造商,医院和外科医生需要解决的因素,以减少这种可避免的临床问题的发生。检索分析是确认大小不匹配的唯一方法,并且可能在国家联合注册管理机构中代表性不足,该机构在修订时记录了修订指示。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号