Patient selection for cementless knee arthroplasty (TKA) remains challenging. 106 consecutive TKAs performed with a cementless porous tantalum tibial component of PS-design were reviewed. Nine failures (8 male/1 female) occurred at a mean 18. months and demonstrated a characteristic failure mode of medial tibial collapse. The mean height of 72.5. inches in the failure group was greater than the 65.8. inches in the well-functioning group (P<. 0.001). The mean weight of 260.4. lb in the failure group was greater than the 204.1. lb in the well-functioning group (P= 0.006). The high early failure rate occurred predominantly in tall, heavy, male patients and may be related to medial tibial overload resulting from increased flexibility of the implant in this patient type.
机译:非骨水泥型膝关节置换术(TKA)的患者选择仍然具有挑战性。回顾了采用PS设计的非骨水泥多孔钽胫骨组件进行的106次连续TKA。平均18个月发生9例衰竭(8例男性/ 1例女性),并表现出胫骨内侧塌陷的特征性故障模式。平均身高为72.5。失败组的英寸数大于65.8。机能良好的组的英寸(P <.0.001)。平均体重为260.4。失败组的磅大于204.1。机能良好组的体重(P = 0.006)。高的早期失败率主要发生在高个,重的男性患者中,并且可能与这种患者类型的植入物的柔性增加所致的内侧胫骨超负荷有关。