Early failure of a roughened surface, precoated femoral component in total hip arthroplasty.

机译:全髋关节置换术中表面粗糙,股骨头预涂的早期失效。

获取原文
获取原文并翻译 | 示例
           

摘要

A retrospective study was performed of a consecutive series examining 98 primary hip arthroplasties between March 1992 and December 1995. A second-generation roughened surface (grit-blasted), polymethyl methacrylate-precoated femoral component (Centralign prosthesis, Zimmer, Inc, Warsaw, IN) was used for all cases. Clinical and radiographic data were available for 84 hips (76 patients). The average duration of follow-up was 35.8 months. There were 10 failures by aseptic mechanical loosening of the femoral component (12%) at an average 30.6 months postoperatively, 9 having been revised (11%). Revision was pending in the remaining patient. The average age of patients at primary arthroplasty for the failed group was 48.2 years compared with 63.4 years for the asymptomatic patients (P =.0008). In all cases, failures occurred by debonding of the femoral prosthesis from the cement mantle. We are not aware of any series using modern implant components and cementing techniques with an early mechanical failure rate of this magnitude.
机译:在1992年3月至1995年12月间,对连续的系列检查进行了回顾性研究,共检查了98例原发性髋关节置换术。第二代粗糙表面(喷砂处理),涂有聚甲基丙烯酸甲酯的股骨组件(Centralign prosthesis,Zimmer,Inc,华沙,印第安纳州) )用于所有情况。临床和影像学数据可用于84髋(76例患者)。平均随访时间为35.8个月。术后平均30.6个月,有10例因股骨组件的无菌机械性松​​动而失败(12%),其中9例因修复而松弛(11%)。其余患者尚待修订。失败组的初次置换患者平均年龄为48.2岁,而无症状患者为63.4岁(P = .0008)。在所有情况下,均因股骨假体与水泥衬套脱开而发生故障。我们尚无任何使用现代植入物组件和固井技术的系列产品,其早期机械失效率如此之高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号