Distal cement mantle thickness with a triangular distal centralizer inserted into the stem tip in cemented total hip arthroplasty.

机译:远端水泥固定器的远端厚度,将三角形远端扶正器插入到全髋关节置换术的茎尖中。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Fifty-seven hips undergoing a primary cemented total hip arthroplasty with use of a triangular distal centralizer inserted into the stem tip were investigated, with a special focus on distal cement mantle thickness. The subjects were 43 women and 3 men. When evaluated on conventional anteroposterior and lateral radiographs, the relative incidence of thin cement mantles in zones 3 or 5 owing to the distal centralizer was 15.8% (9 hips). Unsatisfactory findings were that the distal centralizer and distal part of the stem shifted to the cortex in 7 hips and bent or failed at the inserted site in 2 hips. There is risk of generating a thin cement mantle with use of the triangular centralizer and its insertion into the stem tip.
机译:研究了57例接受了全髋关节置换术的髋关节,并在茎尖插入了三角形远侧扶正器,重点研究了远侧骨水泥的厚度。受试者为43名女性和3名男性。在常规前后位片上进行评估时,由于远端扶正器,在3区或5区水泥薄地幔的相对发生率为15.8%(9髋)。不满意的发现是远端扶正器和茎干的远端部分在7个髋部转移到皮质,并在2个髋部的插入部位弯曲或失败。使用三角形扶正器并将其插入阀杆尖端可能会产生薄的水泥地幔。