首页> 外文期刊>The Journal of arthroplasty >High Failure Rates of the Spectron EF Stem at a Minimum of 10 Year's Follow-up
【24h】

High Failure Rates of the Spectron EF Stem at a Minimum of 10 Year's Follow-up

机译:在至少10年的随访中,Spectron EF阀杆的高故障率

获取原文
获取原文并翻译 | 示例
       

摘要

We prospectively followed 112 hips, undergoing THA with a Spectron EF stem. At mean follow-up of 11.2 years, 21 patients had died. We obtained radiological follow-up in 99% and clinical follow-up in 100% of the surviving 91 hips. Fifty-four percent demonstrated osteolysis in at least one Gruen zone. Twenty-two hips required revision for all causes, with a further five stems radiologically loose. With endpoint being stem revision for aseptic loosening or radiological failure, survivorship at 11 years was 0.783. We believe the addition of a rougher surface finish has contributed to the high levels of osteolysis and stem failure seen with the Spectron EF. Crown Copyright (c) 2014 Published by Elsevier Inc. All rights reserved.
机译:我们前瞻性追踪了112髋,并使用Spectron EF茎进行了THA。平均随访11.2年,有21例患者死亡。在幸存的91例髋关节中,我们进行了99%的放射学随访和100%的临床随访。 54%的患者在至少一个Gruen区表现出骨溶解。所有原因均需翻修22个髋关节,另外5个在放射学上松动的茎。以无菌性松动或放射学衰竭为目的的茎翻修,11年生存率为0.783。我们认为,添加更粗糙的表面光洁度有助于Spectron EF看到高水平的骨溶解和茎干衰竭。官方版权(c)2014,由Elsevier Inc.发行。保留所有权利。